前列腺癌立体定向体部放疗中多叶准直器追踪与治疗床追踪的比较
该研究基于十例前列腺癌患者,在容积旋转调强技术下比较多叶准直器(MLC)追踪与治疗床追踪在立体定向体部放疗(SBRT)中的几何与剂量学表现。几何评估显示,中位过/欠曝光面积从无运动补偿的2.02cm²分别降至MLC追踪的1.00cm²和治疗床追踪的0.45cm²。剂量学上与静态参考最接近的是治疗床追踪,结论为两种追踪均显著提高运动存在时的前列腺SBRT准确性,治疗床追踪略优。
要点
- MLC追踪与治疗床追踪均显著降低前列腺SBRT运动下的过/欠曝光面积。
- 治疗床追踪在几何与剂量学表现上均略优于MLC追踪。
- 无运动补偿时中位过/欠曝光面积为2.02cm²,治疗床追踪降至0.45cm²。
核心结论
- 在常规直线加速器上,MLC追踪与治疗床追踪均显著提高前列腺SBRT在运动存在时的准确性,治疗床追踪表现略优。
研究设计与样本
- 对十例前列腺癌患者,采用容积旋转调强技术制备整合加量的SBRT治疗计划。
- 几何评估将束流等中心处的铅球按五条前列腺运动曲线移动,分别在不补偿、MLC追踪或治疗床追踪下照射并拍摄MV图像评估过/欠曝光区域。
- 剂量学评估将计划施加于剂量学模体,比较有/无运动补偿时靶区与危及器官内的剂量分布。
主要发现
- 中位过/欠曝光面积从无运动补偿的2.02cm²显著降至MLC追踪的1.00cm²和治疗床追踪的0.45cm²。
- 剂量学上与静态参考最接近的结果由治疗床追踪实现。
临床怎么用
- 在前列腺SBRT中可考虑采用MLC追踪或治疗床追踪以减轻运动影响,治疗床追踪的几何与剂量学表现略优。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE AND BACKGROUND: Motion mitigation during prostate stereotactic body radiation therapy (SBRT) ensures optimal target coverage while reducing the risk of overdosage of nearby organs. The geometrical and dosimetrical performance of motion mitigation with the multileaf-collimator (MLC tracking) or the treatment couch (couch tracking) were compared. MATERIAL AND METHODS: For ten prostate patients, SBRT treatment plans with integrated boosts were prepared using volumetric modulated arc technique. For the geometrical evaluation, a lead sphere at the beam isocenter was moved according to five prostate motion curves (i) without mitigation, (ii) with MLC tracking or (iii) with couch tracking. During irradiation, MV images were taken and the over-/underexposed areas were evaluated. For the dosimetrical evaluation, the plans were applied to a dosimetric phantom. Dose distributions with and without mitigation were evaluated inside the target structure and organs at risk. RESULTS: The median over-/underexposed area was reduced significantly from 2.02cm2 without mitigation to 1.00cm2 and 0.45cm2 with MLC and couch tracking. Closest dosimetrical agreement to the static references was achieved with couch tracking. CONCLUSIONS: MLC and couch tracking at a conventional linear accelerator significantly improved the accuracy of prostate SBRT in the presence of motion, whereby couch tracking showed slightly better performance than MLC tracking.
引用
- Ehrbar S, Schmid S, Jöhl A, et al. Comparison of multi-leaf collimator tracking and treatment-couch tracking during stereotactic body radiation therapy of prostate cancer. Radiother Oncol. 2017. PMID: 28935181. doi:10.1016/j.radonc.2017.08.035
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。