前列腺癌多源旋转屏蔽近距离治疗装置
本文提出一种新型多源旋转屏蔽近距离治疗(RSBT)装置,用于在前列腺癌间质插植针中同时精确角度和线性定位部分屏蔽的153Gd源。结论认为该装置可在临床合理时间范围内实现机械可行的尿道保护技术。
要点
- 多源RSBT装置可同步精确控制源的角度和线性位置。
- 相比传统HDR近距离治疗,尿道D0.1cm3降低31%。
核心结论
- 所提出的RSBT递送装置可在临床合理时间范围内实现机械可行的前列腺癌尿道保护治疗技术。
研究设计与样本
- 本研究为装置设计与治疗计划研究,提出多源旋转屏蔽近距离治疗装置并评估其可行性。
主要发现
- RSBT装置使尿道D0.1cm3较传统高剂量率近距离治疗降低31%(尿道表面3mm内剂量梯度体积)。
临床怎么用
- 该装置有望在前列腺癌近距离治疗中降低尿道剂量,同时保持临床可接受的治疗时间。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Our purpose is to present a novel multisource rotating shield brachytherapy (RSBT) apparatus for the simultaneous precise angular and linear positioning of partially shielded 153Gd brachytherapy sources in interstitial needles for the treatment of locally advanced prostate cancer. It is designed to lower the dose to nearby healthy tissues, the urethra in particular, relative to conventional high-dose-rate brachytherapy techniques. METHODS AND MATERIALS: Following needle implantation through the patient template, an angular drive mechanism is docked to the patient template. Each needle is coupled to a multisource afterloader catheter by a connector passing through a shaft. The shafts are rotated about their axes by translating a moving template between 2 stationary templates. The shafts' surfaces and moving template holes are helically threaded with the same pattern such that translation of the moving template causes simultaneous rotation of the shafts. The rotation of each shaft is mechanically transmitted to the catheter-source-shield combination, inside the needles, via several key-keyway pairs. The catheter angles are simultaneously incremented throughout treatment, and only a single 360° rotation of all catheters is needed for a full treatment. For each rotation angle, source depth in each needle is controlled by a multisource afterloader, which is proposed as an array of belt-driven linear actuators, each of which drives a wire that controls catheter depth in a needle. RESULTS: Treatment plans demonstrated that RSBT with the proposed apparatus reduced urethral D0.1cm3 (the minimum dose delivered to the hottest 0.1cm3 of the urethra) below that of conventional high-dose-rate brachytherapy by 31% for urethral dose gradient volume within 3 mm of the urethra surface. Treatment time to deliver 20 Gy with the proposed multisource RSBT apparatus by use of nineteen 62.4-GBq 153Gd sources was 122 minutes. CONCLUSIONS: The proposed RSBT delivery apparatus enables a mechanically feasible urethra-sparing treatment technique for prostate cancer in a clinically reasonable time frame.
引用
- Dadkhah H, Hopfensperger KM, Kim Y, et al. Multisource Rotating Shield Brachytherapy Apparatus for Prostate Cancer. Int J Radiat Oncol Biol Phys. 2017. PMID: 28843372. doi:10.1016/j.ijrobp.2017.06.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。