多数IROC-Houston体模失败案例存在治疗计划系统计算误差

Treatment Planning System Calculation Errors Are Present in Most Imaging and Radiation Oncology Core-Houston Phantom Failures.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2017-04-04

研究利用IROC-Houston测量参考剂量学数据建立独立重算系统,对259例头颈部体模照射进行重算并与机构TPS计算和实测剂量比较。结果显示17%(n=45)头颈部照射存在显著TPS误差,未通过IROC-Houston标准的照射中68%(n=13)存在计算误差。

要点

  • 独立重算系统可识别TPS计算误差并反馈给机构。
  • 17%(n=45)头颈部体模照射存在显著TPS计算误差。
  • 未通过IROC-Houston标准的照射中68%(n=13)有计算误差。

核心结论

  • 近五分之一的机构在IMRT计算中存在TPS误差,凸显TPS中仔细束流建模与计算的必要性。
  • 独立重算系统有助于识别TPS误差的存在并将相关知识反馈给机构。

研究设计与样本

  • 基于IROC-Houston常见直线加速器类别的测量参考剂量学数据,调试并定制了独立重算系统。
  • 使用该系统重算了259例头颈部体模照射,并将重算结果与机构TPS计算分别同实测剂量比较。

主要发现

  • 平均而言,参考重算系统的准确度优于机构TPS。
  • 17%(n=45)的头颈部照射存在显著TPS误差。
  • 68%(n=13)未通过IROC-Houston标准的照射存在计算误差。
  • 重算比机构TPS平均准确2%或单点准确3%时,该照射被标记为存在「显著」机构计算误差。

关键数值与推荐

  • 显著机构计算误差判定标准:重算比机构TPS平均准确2%或单点准确3%。
  • 17%(n=45)头颈部照射存在显著TPS误差。
  • 68%(n=13)未通过IROC-Houston标准的照射存在计算误差。

临床怎么用

  • 对未通过IROC-Houston端到端体模测试的机构,可用独立重算系统定位TPS计算误差。
  • 应重视TPS束流建模与剂量计算的准确性,并结合独立重算识别和纠正问题。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The anthropomorphic phantom program at the Houston branch of the Imaging and Radiation Oncology Core (IROC-Houston) is an end-to-end test that can be used to determine whether an institution can accurately model, calculate, and deliver an intensity modulated radiation therapy dose distribution. Currently, institutions that do not meet IROC-Houston's criteria have no specific information with which to identify and correct problems. In the present study, an independent recalculation system was developed to identify treatment planning system (TPS) calculation errors. METHODS AND MATERIALS: A recalculation system was commissioned and customized using IROC-Houston measurement reference dosimetry data for common linear accelerator classes. Using this system, 259 head and neck phantom irradiations were recalculated. Both the recalculation and the institution's TPS calculation were compared with the delivered dose that was measured. In cases in which the recalculation was statistically more accurate by 2% on average or 3% at a single measurement location than was the institution's TPS, the irradiation was flagged as having a "considerable" institutional calculation error. The error rates were also examined according to the linear accelerator vendor and delivery technique. RESULTS: Surprisingly, on average, the reference recalculation system had better accuracy than the institution's TPS. Considerable TPS errors were found in 17% (n=45) of the head and neck irradiations. Also, 68% (n=13) of the irradiations that failed to meet the IROC-Houston criteria were found to have calculation errors. CONCLUSIONS: Nearly 1 in 5 institutions were found to have TPS errors in their intensity modulated radiation therapy calculations, highlighting the need for careful beam modeling and calculation in the TPS. An independent recalculation system can help identify the presence of TPS errors and pass on the knowledge to the institution.