大型头颈肿瘤中心常规IMRT质量保证结果
该研究回顾性分析前瞻性数据库中的根治性头颈肿瘤患者,评估常规IMRT质量保证(QA)流程中重大修改的实施比例是否超过10%。共548例患者纳入,92.1%依从性,14.8%患者实施至少一项重大修改,仅IMRT治疗与重大修改可能性相关(P=.001),85%的QA耗时少于20分钟。
要点
- 常规IMRT QA中14.8%患者实施至少一项重大修改,超过预设10%的终点。
- 仅IMRT治疗与重大修改可能性相关(P=.001)。
- 85%的IMRT QA流程耗时少于20分钟。
核心结论
- 常规头颈IMRT QA项目导致大量重大修改的实施,高依从性表明对其临床价值的认可。
研究设计与样本
- 回顾性分析前瞻性数据库,纳入2012年5月21日至2014年5月21日根治性治疗的头颈肿瘤患者,共548例进入分析,依从性92.1%。
主要发现
- 115例(21%)患者实施了次要修改建议,13例(2.4%)未实施,324例(59.1%)无修改建议。
- 仅IMRT治疗与重大修改可能性相关(P=.001)。
- 85%的病例IMRT QA耗时少于20分钟。
关键数值与推荐
- 主要终点:常规IMRT QA流程中重大修改实施比例是否超过10%。
- 推荐修改预定义为「重大修改」或「次要修改」。
临床怎么用
- 常规头颈IMRT QA项目可发现并实施大量重大修改,且耗时较短,支持其在临床实践中的价值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The primary endpoint was to ascertain whether the percentage of major changes implemented from our routine intensity modulated radiation therapy (IMRT) quality assurance (QA) process was more than 10%. The secondary endpoints were to document the percentage of minor changes, ascertain the time taken to perform the IMRT QA process, evaluate potential predictors for major changes, and ascertain the perceived value of the program by the compliance of radiation oncologists (ROs) treating head and neck cancer (HNC). METHODS AND MATERIALS: This was a retrospective analysis of a prospective database for all radically treated HNC patients. Recommended changes were predefined with criteria as either "major changes" or "minor changes." RESULTS: Of 595 patients treated radically between May 21, 2012, and May 21, 2014, 548 patients were entered, giving a compliance of 92.1%. The vast majority were treated with IMRT (470/548, 89%), 49.3% treated definitively and 50.7% treated adjuvantly; overall, 63% had stage IV disease. Eighty-one patients (14.8%) had 1 or more major changes recommended and implemented, and 21 patients (3.8%) had major changes recommended but not implemented because of a lack of consensus. Of minor recommendations, in 115 patients (21%) they were implemented and in 13 patients (2.4%) they were not implemented. No changes were recommended in the remaining 324 patients (59.1%). The only factor correlating with the likelihood of a major change was treatment with IMRT (P=.001). In the vast majority of cases (85%), the time taken for IMRT QA was less than 20 minutes. CONCLUSIONS: Our routine HNC IMRT QA program resulted in the implementation of a significant number of major changes. The high compliance with the program is indicative of the belief in its clinical value. This study can inform discussion on optimal RT QA processes in routine clinical practice.
引用
- Amarasena I, Herschtal A, D'Costa I, et al. Outcomes of Routine Intensity Modulated Radiation Therapy Quality Assurance in a Large Head and Neck Cancer Center. Int J Radiat Oncol Biol Phys. 2017. PMID: 28581395. doi:10.1016/j.ijrobp.2017.02.215
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。