局部晚期肺癌视觉引导深吸气屏气放疗的靶区位置不确定性
本研究在9例局部晚期非小细胞肺癌(NSCLC)患者视觉引导深吸气屏气(DIBH)放疗中,用垂直透视电影和图像追踪算法评估靶区位置不确定性。结果显示DIBH较自由呼吸(FB)标记点位移平均减少2-6mm,屏气内系统误差与随机误差小,屏气间平均偏移小,但靶区运动未完全消除。
要点
- DIBH较FB在前后、左右、头脚方向标记点位移平均减少2-6mm。
- DIBH屏气内系统误差与随机误差分别为0.5/0.3mm(AP)、0.5/0.3mm(LR)、0.8/0.4mm(CC)。
- 屏气间平均偏移为-0.3mm(AP)、-0.2mm(LR)、-0.2mm(CC)。
核心结论
- 视觉引导DIBH放疗中NSCLC肿瘤和淋巴结的屏气内与屏气间不确定性均较小。
- 视觉引导DIBH可显著减少但未完全消除靶区运动。
研究设计与样本
- 对9例局部晚期NSCLC患者视觉引导DIBH放疗疗程中采集自由呼吸和DIBH垂直透视电影。
- 患者纵隔淋巴结和原发肿瘤内注射液体标记物,采用基于图像的追踪算法分析位移、系统误差、随机误差及屏气间位置不确定性。
主要发现
- DIBH较FB标记点位移在前后、左右和头脚方向平均减少2-6mm。
- DIBH期间淋巴结运动源于心脏运动。
- DIBH系统误差(所有平均标记位置的标准差)和随机误差(屏气内标准差的均方根)分别为0.5和0.3mm(AP)、0.5和0.3mm(LR)、0.8和0.4mm(CC)。
- 屏气间平均偏移为-0.3mm(AP)、-0.2mm(LR)、-0.2mm(CC)。
关键数值与推荐
- 系统误差与随机误差:AP 0.5/0.3mm,LR 0.5/0.3mm,CC 0.8/0.4mm。
- 屏气间平均偏移:AP -0.3mm,LR -0.2mm,CC -0.2mm。
临床怎么用
- 视觉引导DIBH可用于减少局部晚期NSCLC放疗靶区运动,但需注意残余不确定性尤其是头脚方向。
- 纵隔淋巴结运动受心脏运动影响,规划时需考虑。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The purpose of this study was to estimate the uncertainty in voluntary deep-inspiration breath-hold (DIBH) radiotherapy for locally advanced non-small cell lung cancer (NSCLC) patients. METHODS: Perpendicular fluoroscopic movies were acquired in free breathing (FB) and DIBH during a course of visually guided DIBH radiotherapy of nine patients with NSCLC. Patients had liquid markers injected in mediastinal lymph nodes and primary tumours. Excursion, systematic- and random errors, and inter-breath-hold position uncertainty were investigated using an image based tracking algorithm. RESULTS: A mean reduction of 2-6mm in marker excursion in DIBH versus FB was seen in the anterior-posterior (AP), left-right (LR) and cranio-caudal (CC) directions. Lymph node motion during DIBH originated from cardiac motion. The systematic- (standard deviation (SD) of all the mean marker positions) and random errors (root-mean-square of the intra-BH SD) during DIBH were 0.5 and 0.3mm (AP), 0.5 and 0.3mm (LR), 0.8 and 0.4mm (CC), respectively. The mean inter-breath-hold shifts were -0.3mm (AP), -0.2mm (LR), and -0.2mm (CC). CONCLUSION: Intra- and inter-breath-hold uncertainty of tumours and lymph nodes were small in visually guided breath-hold radiotherapy of NSCLC. Target motion could be substantially reduced, but not eliminated, using visually guided DIBH.