OSLD用于适形与调强放疗的在体剂量学:两年多中心队列研究

In vivo dosimetry with optically stimulated luminescent dosimeters for conformal and intensity-modulated radiation therapy: A 2-year multicenter cohort study.
PubMed 文献 · Pract Radiat Oncol · 2016-09-07

该研究基于4台直线加速器、为期2年的每周在体剂量学(IVD)测量,分析10,224次光释光剂量计(OSLD)测量数据,按模态、技术、放置位置和解剖部位分组,用高斯模型拟合。结果显示总体均值±标准差为0.3±10.3%,电子线精度(6.9%)显著优于光子线(10.5%),不同技术精度差异显著(P<.0001),野中野最低(σ=7.2%),IMRT和VMAT最高(σ=11.9%和13.4%)。结论认为±3%的准确度可实现,精度范围6.9%-13.4%,标准化放置位置可能改善IVD精度。

要点

  • 总体OSLD在体剂量测量均值±标准差为0.3±10.3%,准确度±3%可实现。
  • 电子线精度(6.9%)显著优于光子线(10.5%);野中野精度最低(σ=7.2%),IMRT和VMAT最高(σ=11.9%和13.4%)。
  • 头颈部位准确度偏差为-3.7%,乳腺精度最佳而头颈最差(范围7.3%-13.0%)。

核心结论

  • OSLD在体剂量学的准确度可达±3%,精度随模态、技术和治疗部位在6.9%至13.4%间变化。
  • 简单、标准化的放置位置可能改善在体剂量学精度,研究结果有助于制定患者特异性容差。

研究设计与样本

  • 多中心队列研究,收集4台直线加速器为期2年的每周在体剂量学测量数据。
  • 经离群值分析后纳入10,224次测量,按模态、治疗技术、放置位置和解剖部位分组。
  • 采用高斯函数建模,最小二乘法拟合,以决定系数评估拟合优度,方差分析和Levene检验比较组间差异(α=0.05)。

主要发现

  • 总体均值±标准差为0.3±10.3%;电子线精度6.9%显著优于光子线10.5%。
  • 不同治疗技术精度差异显著(P<.0001):野中野最低(σ=7.2%),IMRT和VMAT最高(σ=11.9%和13.4%)。
  • 10个治疗部位中6个模型拟合优度大于0.90,其准确度在±3%以内,但头颈部位均值偏差为-3.7%。
  • 精度随治疗部位变化(范围7.3%-13.0%),乳腺最佳、头颈最差。
  • 位于主射束角度中心轴的放置位置精度(σ=8.5%)优于其他位置(范围10.5%-11.4%)。

关键数值与推荐

  • 总体OSLD在体剂量测量均值±标准差为0.3±10.3%。
  • 电子线测量精度6.9%,光子线10.5%。
  • 野中野、IMRT、VMAT的精度分别为σ=7.2%、11.9%、13.4%(P<.0001)。
  • 头颈部位准确度均值偏差为-3.7%;部位精度范围7.3%-13.0%。
  • 中心轴放置精度σ=8.5%,其他位置范围10.5%-11.4%。

临床怎么用

  • OSLD在体剂量学可用于IMRT和VMAT等现代放疗技术的常规质量保证,准确度±3%可实现。
  • 解读测量结果时应考虑模态、技术和治疗部位对精度的显著影响,避免统一容差。
  • 放置于主射束角度中心轴等标准化位置可能提高在体剂量学精度。
  • 研究数据可为制定基于OSLD的患者特异性容差提供参考。

局限性

  • 原文摘要未明确报告局限性内容。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Optically stimulated luminescent dosimeters (OSLDs) are utilized for in vivo dosimetry (IVD) of modern radiation therapy techniques such as intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT). Dosimetric precision achieved with conventional techniques may not be attainable. In this work, we measured accuracy and precision for a large sample of clinical OSLD-based IVD measurements. METHODS AND MATERIALS: Weekly IVD measurements were collected from 4 linear accelerators for 2 years and were expressed as percent differences from planned doses. After outlier analysis, 10,224 measurements were grouped in the following way: overall, modality (photons, electrons), treatment technique (3-dimensional [3D] conformal, field-in-field intensity modulation, inverse-planned IMRT, and VMAT), placement location (gantry angle, cardinality, and central axis positioning), and anatomical site (prostate, breast, head and neck, pelvis, lung, rectum and anus, brain, abdomen, esophagus, and bladder). Distributions were modeled via a Gaussian function. Fitting was performed with least squares, and goodness-of-fit was assessed with the coefficient of determination. Model means (μ) and standard deviations (σ) were calculated. Sample means and variances were compared for statistical significance by analysis of variance and the Levene tests (α = 0.05). RESULTS: Overall, μ ± σ was 0.3 ± 10.3%. Precision for electron measurements (6.9%) was significantly better than for photons (10.5%). Precision varied significantly among treatment techniques (P < .0001) with field-in-field lowest (σ = 7.2%) and IMRT and VMAT highest (σ = 11.9% and 13.4%, respectively). Treatment site models with goodness-of-fit greater than 0.90 (6 of 10) yielded accuracy within ±3%, except for head and neck (μ = -3.7%). Precision varied with treatment site (range, 7.3%-13.0%), with breast and head and neck yielding the best and worst precision, respectively. Placement on the central axis of cardinal gantry angles yielded more precise results (σ = 8.5%) compared with other locations (range, 10.5%-11.4%). CONCLUSIONS: Accuracy of ±3% was achievable. Precision ranged from 6.9% to 13.4% depending on modality, technique, and treatment site. Simple, standardized locations may improve IVD precision. These findings may aid development of patient-specific tolerances for OSLD-based IVD.