经会阴超声与锥形束CT在前列腺癌外照射放疗中患者摆位验证的初步结果

Initial results for patient setup verification using transperineal ultrasound and cone beam CT in external beam radiation therapy of prostate cancer.
PubMed 文献 · Radiat Oncol · 2016-11-08

该研究对10例接受外照射放疗的前列腺癌患者,连续采集150次经会阴超声(TPUS)与锥形束CT(CBCT)扫描,比较两种模态检测的摆位误差。结果显示TPUS与CBCT在三个方向上的平均差异分别为侧向(-2.7±2.3)mm、垂直向(3.0±2.4)mm和纵向(3.2±2.7)mm,平均欧几里得差异为(6.0±3.1)mm;58%的治疗会话中两者差异至少5 mm。作者认为CBCT摆位验证联合TPUS分次内运动监测可结合两种成像模态的优势。

要点

  • TPUS与CBCT检测的摆位误差平均欧几里得差异为(6.0±3.1)mm。
  • 58%的监测治疗会话中TPUS与CBCT差异至少5 mm。
  • 两种模态摆位误差相关性较弱,但联合使用可兼顾摆位验证与分次内监测。

核心结论

  • TPUS与CBCT之间的摆位差异平均为6 mm。
  • 虽然两种成像模态确定的摆位误差相关性较弱,但CBCT摆位验证联合TPUS分次内运动监测可利用两种模态的优势。

研究设计与样本

  • 对10例接受外照射放疗的前列腺癌患者,使用经会阴超声(TPUS)和锥形束CT(CBCT)进行摆位验证。
  • 共连续采集150次超声和CBCT扫描,分析摆位误差,并分别评估两种模态在各维度上的摆位误差偏差。

主要发现

  • TPUS与CBCT在侧向、垂直向和纵向的平均差异分别为(-2.7±2.3)mm、(3.0±2.4)mm和(3.2±2.7)mm。
  • TPUS与CBCT的平均欧几里得差异为(6.0±3.1)mm。
  • 两种成像模态之间观察到的差异高达19.2 mm。
  • 58%的监测治疗会话中TPUS与CBCT差异至少5 mm。

临床怎么用

  • CBCT可用于摆位验证,TPUS可用于分次内运动监测,两者联合可发挥各自优势。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Evaluation of set up error detection by a transperineal ultrasound in comparison with a cone beam CT (CBCT) based system in external beam radiation therapy (EBRT) of prostate cancer. METHODS: Setup verification was performed with transperineal ultrasound (TPUS) and CBCT for 10 patients treated with EBRT for prostate cancer. In total, 150 ultrasound and CBCT scans were acquired in rapid succession and analyzed for setup errors. The deviation between setup errors of the two modalities was evaluated separately for each dimension. RESULTS: A moderate correlation in lateral, vertical and longitudinal direction was observed comparing the setup errors. Mean differences between TPUS and CBCT were (-2.7 ± 2.3) mm, (3.0 ± 2.4) mm and (3.2 ± 2.7) mm in lateral, vertical and longitudinal direction, respectively. The mean Euclidean difference between TPUS and CBCT was (6.0 ± 3.1) mm. Differences up to 19.2 mm were observed between the two imaging modalities. Discrepancies between TPUS and CBCT of at least 5 mm occurred in 58 % of monitored treatment sessions. CONCLUSION: Setup differences between TPUS and CBCT are 6 mm on average. Although the correlation of the setup errors determined by the two different image modalities is rather week, the combination of setup verification by CBCT and intrafraction motion monitoring by TPUS imaging can use the benefits of both imaging modalities.