文献速递放疗物理与技术姑息放疗与转移瘤

资源有限地区放射肿瘤科事件学习系统的实施:风险分析与安全文化

Implementation of a Radiation Oncology Incident Learning System in a Limited Resource Setting: Risk Analysis and Safety Culture.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2026-08-21

PURPOSE: Few reports address the use of incident learning systems (ILS) in low- and middle-income countries (LMICs). We hypothesize that ILS can support quality improvement (QI) needs in an LMIC clinic, and that a commitment to QI can overcome perceived barriers to ILS uptake.…

要点

  • 治疗计划是最常见错误来源,全面物理计划检查可预防多数错误。
  • 最高风险失效模式与OIS相关的治疗实施错误有关。

核心结论

  • ILS报告提示质量改进需求集中在治疗计划、技术计划审查以及OIS实施。
  • 对质量改进的承诺可克服ILS使用的感知障碍。

研究设计与样本

  • 在撒哈拉以南非洲一家放射肿瘤科诊所实施基于国际原子能机构分类法的ILS,并开展初步安全文化调查。
  • 七个月后按类型、区域、角色和来源分析ILS报告,使用QCQ确定最可能预防或检测错误的QA干预,并由多学科团队用FMEA对23个代表性失效模式按RPN排序。

主要发现

  • 调查显示报告者担心声誉或报告同事的后果是事件报告的障碍。
  • 多数报告从直线加速器控制室录入(41份,48.8%),但治疗计划是最常见错误来源(28份,33.3%)。
  • 最常识别的可预防错误的QA干预为全面物理计划检查(29份报告)。
  • 最高RPN失效模式与OIS相关的治疗实施错误有关。

临床怎么用

  • 在资源有限地区可实施ILS以支持质量改进。
  • 应重点关注治疗计划和技术计划审查的质量控制。
  • 需关注OIS实施相关的治疗实施错误风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Few reports address the use of incident learning systems (ILS) in low- and middle-income countries (LMICs). We hypothesize that ILS can support quality improvement (QI) needs in an LMIC clinic, and that a commitment to QI can overcome perceived barriers to ILS uptake. METHODS: A preliminary survey assessing safety culture and perceived barriers to ILS use was distributed to all clinical staff at a radiation oncology clinic in sub-Saharan Africa. An ILS was implemented based on the International Atomic Energy Agency taxonomy. After seven months, ILS reports were analyzed by type, clinic area, clinical role, and origin. Quality control quantification (QCQ) was used to determine the QA intervention most likely to prevent or detect each error. An interdisciplinary team of clinicians used failure modes and effects analysis (FMEA) to rank 23 representative failure modes by risk priority number (RPN). RESULTS: Survey responses indicated that "concern on the part of the reporter about their reputation or the effects of reporting a colleague" was a barrier to incident reporting. Despite this, the ILS collected eighty-four reports throughout the seven-month accrual period. Most were classified as near-misses (32, 38%). Although most reports were entered from the LINAC control rooms (41, 48.8%), treatment planning was the most common origin of errors (28, 33.3%). The most frequently identified QA intervention that would have prevented errors was a comprehensive physics plan check (29 reports). The highest RPN failure mode was associated with a treatment delivery error related to the oncology information system (OIS). CONCLUSION: ILS reports indicate that QI needs center on treatment planning and technical plan review, as well as OIS implementation. This study, one of the few to explore the use of ILS in the LMIC setting, illustrates how a commitment to QI can overcome perceived barriers to ILS use.

来源:Int J Radiat Oncol Biol Phys,查看原文。版权归原作者所有。

引用

  1. Colbert C, Komakech I, Kavuma A, et al. Implementation of a Radiation Oncology Incident Learning System in a Limited Resource Setting: Risk Analysis and Safety Culture. Int J Radiat Oncol Biol Phys. 2026. PMID: 42628810. doi:10.1016/j.ijrobp.2026.07.067

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。