儿童肿瘤协作组物理委员会报告:影像与放射肿瘤学核心现场认证及病例审查结果

A Children's Oncology Group Physics Committee Report on Imaging and Radiation Oncology Core Site Credentialing and Patient Case Review Results.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2026-07-24

该研究由儿童肿瘤协作组(COG)物理委员会代表开展,回顾分析2013-2023年10年间已关闭的COG方案,评估体模认证表现并与病例审查偏差比较。研究发现认证与病例审查互为补充,能识别独特失效模式,常在影响入组患者前发现错误;尽管最终方案偏差率低,但病例审查和认证过程识别出许多错误,凸显临床试验放疗数据质量保证的必要性。

要点

  • 体模认证可发现束流建模或患者摆位流程错误,病例审查可发现解剖勾画、处方、计划技术和剂量覆盖错误。
  • 认证与病例审查互为补充,常在错误影响入组患者前识别独特失效模式。
  • 尽管COG方案最终偏差率低,但质量保证检查仍不可缺。

核心结论

  • 尽管COG方案最终方案偏差率低,但病例审查和认证过程识别出许多错误,凸显临床试验放疗数据质量保证检查的持续必要性。

研究设计与样本

  • 由COG物理委员会代表开展,回顾分析10年期间(2013-2023)已关闭的COG方案。
  • 评估COG机构的体模认证表现并与病例审查偏差比较以确定趋势,汇总体模照射失败并审查常见失效模式,对病例审查的方案偏差进行分类并描述模式。

主要发现

  • 体模认证与病例审查在临床试验质量保证中互为补充,识别独特失效模式,常在影响入组患者前发现。
  • 体模认证识别束流建模或患者摆位流程错误,病例审查识别解剖勾画、处方、治疗计划技术和剂量覆盖错误。

临床怎么用

  • COG方案根据允许的治疗技术要求特定现场认证活动,并采用方案特定治疗计划指南定义肿瘤覆盖和正常组织保护的可接受限值、临床必要时的变异及方案偏差。
  • 认证活动确保放疗设备和交付符合标准,个体患者病例审查确保每个方案中靶区勾画、肿瘤覆盖和正常组织避让的一致性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The Imaging and Radiation Oncology Core works with the Children's Oncology Group (COG) to provide institutional credentialing and patient radiation therapy plan review for sites that enroll patients on COG clinical trials. COG protocols that include radiation therapy require specific site credentialing activities depending on the treatment technique that is allowed in the protocol. In addition, there are protocol-specific treatment planning guidelines for tumor coverage and normal tissue sparing that define not only protocol acceptable limits but also variations in limits if deemed clinically necessary, as well as protocol deviations. With many institutions participating in COG protocols, credentialing activities ensure standards for radiation therapy equipment and delivery are met and individual patient case reviews ensure consistency in target delineation, tumor coverage, and normal tissue avoidance for each protocol. The purpose of this work is to provide a descriptive analysis of retrospective credentialing and case review data for COG clinical trials. METHODS AND MATERIALS: This work was conducted by representatives of the COG Physics Committee and aimed to characterize protocol deviations of recently closed COG trials. This analysis included closed COG protocols from a 10-year period (2013-2023). Phantom credentialing performance of COG institutions was assessed and compared with case review deviations to determine trends. Credentialing phantom irradiation failures were collated and reviewed for common failure modes. From patient case reviews, protocol deviations were categorized, and patterns of protocol deviations are described. RESULTS: This study found that credentialing and case reviews in clinical trial quality assurance were complementary, identifying unique failure modes, often before they affected enrolled patients. For example, phantom credentialing identified errors in beam modeling or patient setup processes, whereas case reviews identified errors in anatomical contouring, prescriptions, treatment planning techniques, and dose coverage. CONCLUSIONS: Although the final rate of protocol deviations was low for COG protocols, many errors were identified through the case review and credentialing processes, highlighting the continued need for quality assurance checks of radiation therapy data in clinical trials.

来源:Int J Radiat Oncol Biol Phys,查看原文。版权归原作者所有。