优化大野乳房切除术后质子透射FLASH放疗

Optimizing Transmission FLASH Radiation Therapy for Large-Field Postmastectomy Breast Treatment.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2026-05-08

本研究对5例左侧乳房切除术后乳腺癌患者的质子透射束计划进行回顾性再计划,评估扫描速度、束流配置和剂量率模型对FLASH效应的影响。结果表明FLASH效应高度依赖扫描轨迹、扫描速度和模型选择,采用遗传算法优化束斑序列可进一步改善健康组织的剂量率分布。

要点

  • 垂直扫描速度从10增至20 mm/ms可提高CTV和皮肤的FLASH效应。
  • 分野照射通过增加纵向列间时间距离改善FLASH效应。
  • 遗传算法路径优化可缩短扫描时间并获得与分野照射相当的FLASH效应。

核心结论

  • FLASH效应高度依赖于扫描轨迹、扫描速度和模型选择,采用遗传算法等路径最小化方法优化束斑投照可进一步改善所有场景下健康体素的剂量率分布。

研究设计与样本

  • 回顾性再计划研究,纳入5例左侧乳房切除术后乳腺癌患者,处方剂量32 Gy/5次,采用单能249 MeV切线透射束并辅以临床垂直束均整剂量。

主要发现

  • 垂直扫描速度从10增至20 mm/ms,CTV的FLASH效应分别增加22%(ADR模型)和12%(FEM模型),皮肤从41.4%升至58.8%(ADR)和8.4%升至13.1%(FEM)。
  • 分野照射改善纵向列间时间距离,ADR模型下CTV平均剂量(FLASH校正)最多降低9.2 Gy。
  • 基于遗传算法的优化缩短扫描时间,CTV平均剂量降低7.87 Gy(ADR GA),皮肤平均剂量降低2至3 Gy。

关键数值与推荐

  • FLASH效应量化标准为体素满足剂量≥4 Gy且剂量率≥40 Gy/s的百分比。
  • 对满足FLASH标准的体素应用0.67的剂量调整因子计算FLASH加权剂量指标,肿瘤细胞物理剂量保持不变。

临床怎么用

  • 在质子透射FLASH计划设计中应综合考虑扫描速度、束流配置和剂量率模型选择,并可利用遗传算法等路径优化方法改善健康组织剂量率分布。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: We investigated the effect of scanning speed, beam configuration, and dose-rate modeling on the FLASH effect in postmastectomy proton transmission beams planning and evaluated the potential of spot scanning path optimization for enhancing the FLASH effect. METHODS AND MATERIALS: Five patients with left-sided postmastectomy breast cancer (32 Gy/5 fractions) were retrospectively replanned with single-energy (249 MeV) tangential transmission beams, supplemented by a clinical en face beam for dose homogenization. FLASH evaluation employed 2 models: Krieger's FLASH effectiveness model (FEM) and Folkerts' average dose-rate (ADR) framework. Plans were simulated under conventional pencil beam scanning, split-field, and optimized spot sequences (using genetic algorithm [GA]), with vertical scan speeds varied from 10 to 20 mm/ms. FLASH effect in normal tissues was quantified by the percentage of voxels meeting the threshold (≥4 Gy at ≥40 Gy/s). A dose adjustment factor of 0.67 was applied to voxels meeting FLASH criteria to compute FLASH-weighted dose metrics in normal tissues within the chest wall target, whereas the physical dose to tumor cells remained unchanged. RESULTS: The FLASH effect showed high sensitivity to scanning patterns and model selection. Increasing vertical scan speed from 10 to 20 mm/ms increased the FLASH in clinical target volume (CTV) by 22% (ADR) and 12% (FEM), whereas in skin, it rose from 41.4% to 58.8% (ADR) and 8.4% to 13.1% (FEM). Split-field delivery improved the temporal distance between the vertical columns of the spot scanning pattern, yielding a superior FLASH effect, which is up to a 9.2 Gy reduction in CTV (FLASH-corrected) Dmean(mean dose) with the ADR model. GA-based optimization shortened scan time and provided FLASH comparable with split-field delivery, with a CTV Dmean reduction of 7.87 Gy (ADR GA), with skin Dmean reductions of 2 to 3 Gy. CONCLUSION: This study demonstrates that FLASH outcomes are highly sensitive to scanning trajectory, scan speed, and model selection. Beyond these parameters, optimizing spot delivery using a path minimizer, such as GA, can further improve the dose-rate distribution in healthy voxels across all scenarios.