头颈癌CBCT引导在线自适应放疗的分次内运动与 margin 评估
本研究回顾性分析24例头颈鳞癌患者接受CBCT引导每日在线自适应放疗的166个分次,评估分次内运动所需的margin。结果显示计划CBCT至照射结束平均33分钟,1 mm轴向+2 mm头脚方向margin下所有靶区几何重叠>99%且满足剂量覆盖目标,提示每日oART可实现稳健覆盖并支持更小PTV margin。
要点
- 计划CBCT至照射结束平均33分钟,分次内运动margin需求小。
- 1 mm轴向+2 mm头脚方向margin下所有靶区几何重叠>99%。
- 各靶区剂量覆盖V100≥95%和V95≥99%目标均满足。
核心结论
- 尽管自适应流程耗时,每日oART以最小分次内运动margin实现稳健的几何与剂量覆盖,支持头颈鳞癌使用更小PTV margin。
研究设计与样本
- 回顾性分析24例接受CBCT引导每日在线自适应放疗的头颈鳞癌患者,来自前瞻性临床试验,每周评估1个分次,共166个分次。
主要发现
- 计划CBCT至照射结束平均33分钟(绝对范围25-58分钟)。
- 1 mm轴向+2 mm头脚方向margin下所有金标准靶区几何重叠>99%。
- 原发GTV-70几何覆盖99.7%、V100=96.9%;淋巴结GTV-70几何覆盖99.0%、V100=96.6%。
- CTV-63几何覆盖99.4%、V100=98.7%;可疑淋巴结>17 mm的GTVn-63几何覆盖99.1%、V100=96.8%。
- 可疑淋巴结≤17 mm的GTVn-56几何覆盖99.6%、V100=99.0%。
关键数值与推荐
- 剂量覆盖目标定义为V100%≥95%且V95%≥99%。
- 评估的margin为径向1、2、3 mm加头脚方向2 mm(CBCT层厚)。
- 原发GTV-70、淋巴结GTV-70、CTV-63、GTVn-63、GTVn-56的几何覆盖均>99%且满足剂量覆盖目标。
临床怎么用
- CBCT引导每日在线自适应放疗在头颈鳞癌中可考虑使用1 mm轴向+2 mm头脚方向的分次内运动margin。
- 自适应流程时间需关注,本研究中计划CBCT至照射结束平均33分钟。
局限性
- 回顾性分析,样本量24例、166个分次,每周仅评估1个分次。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Daily online adaptive radiation therapy (oART) for head and neck squamous cell carcinoma (HNSCC) holds the potential to holds the potential to dramatically reduce margins accounting for setup uncertainty. However, careful margin assessment is critical due to the additional time required for real-time contouring and planning, which may lead to consequential intrafractional motion. This study aimed to quantify the intrafraction motion margin required for cone beam computed tomography (CBCT) guided oART in HNSCC. METHODS AND MATERIALS: This retrospective analysis included 24 patients with HNSCC who were treated with CBCT guided daily oART as part of a prospective clinical trial. Eligible patients were treated for oropharyngeal, laryngeal, and hypopharyngeal HNSCC. One fraction was assessed per week for each patient. For each of the oART sessions, a preplanning CBCT and a near posttreatment (PT) CBCT, acquired before the last field delivery, were obtained. "Gold standard" (GS) gross tumor volume (GTV) and clinical target volume (CTV) were retrospectively contoured on both CBCTs without time constraints. Radial margins of 1, 2, and 3 mm plus a craniocaudal margin of 2 mm (CBCT slice thickness) were applied to the planning GS contours to assess geometric overlap with PT GS contours. Dosimetric coverage (V100% ≥95% and V95% ≥99%) of the adapted plan generated with preplanning CBCT GS contour on the PT-CBCT GS contour was evaluated. Workflow times, including the interval from planning CBCT to final beam delivery, were recorded. RESULTS: A total of 166 fractions were evaluated, with an average time of 33 minutes (absolute range, 25-58 minutes) between planning CBCT and delivery completion. Geometric overlap and dosimetric coverage analysis showed that a 1-mm axial and 2-mm superior-inferior margin achieved >99% geometric overlap for all GS targets and met dosimetric coverage goals: primary GTV receiving 70 Gy (GTVp-70; 99.7% geometric coverage, V100 = 96.9%), nodal GTV receiving 70 Gy (GTVn-70; 99.0%, 96.6%), primary CTV receiving 63 Gy (CTV-63; 99.4%, 98.7%), nodal GTV receiving 63 Gy for suspicious nodes >17 mm (GTVn-63; 99.1%, 96.8%), and nodal GTV receiving 56 Gy for suspicious nodes ≤17 mm (GTVn-56; 99.6%, 99.0%). CONCLUSIONS: Despite the time required for the adaptive process, daily oART achieves robust geometric and dosimetric coverage with minimal intrafraction motion margin. This stability suggests that daily oART for HNSCC can significantly facilitate the use of a smaller planning target volume margin.
引用
- Liao CY, Domal S, Yarlagadda P, et al. Intrafractional Motion and Margin Assessment for Cone Beam Computed Tomography Guided Online Adaptive Radiation Therapy in Head and Neck Cancer. Adv Radiat Oncol. 2026. PMID: 41890534. doi:10.1016/j.adro.2026.102013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。