光子计数CT用于胰腺癌光子剂量计算的准确性:基于虚拟单能与电子密度图像的比较研究
该概念验证研究纳入29例接受腹部增强光子计数CT(PCCT)扫描的患者,比较70 keV虚拟单能图像(VMI)与电子密度(Rho)图像在胰腺放疗剂量计算中的差异。结果显示两种图像DVH指标差异多在1%以内(最大-1.60%),三维伽马分析通过率多数>98%,表明PCCT可安全用于剂量计算。
要点
- 70 keV VMI与Rho图像剂量计算一致性良好,DVH差异多在1%以内。
- 三维伽马分析(1 mm/1%)通过率多数>98%,等剂量分布相似。
- Rho图像可抑制碘对比剂增强,减少不必要的不确定性。
核心结论
- 70 keV VMI与Rho图像在胰腺放疗剂量计算中具有良好一致性,PCCT可安全用于剂量计算并具一定优势。
研究设计与样本
- 概念验证研究,纳入29例接受腹部增强PCCT扫描的患者,经机构审查委员会批准。
- 比较70 keV VMI、Rho图像及默认HLUT三种剂量计算方式,采用9野IMRT同步整合推量25/33 Gy计划。
主要发现
- 70 keV VMI与Rho图像DVH指标百分比差异多在1%以内,最大差异为-1.60%。
- 多数病例三维伽马分析(1 mm/1%)通过率>98%,差异最大病例平均DVH差异-1.32%、通过率94.82%。
- 使用Eclipse默认HLUT的剂量计算DVH差异在1.5%范围内。
关键数值与推荐
- DVH指标百分比差异多在1%以内,最大-1.60%。
- 三维伽马分析(1 mm/1%)通过率多数>98%,最低94.82%。
- 默认HLUT剂量计算DVH差异在1.5%范围内。
临床怎么用
- PCCT的70 keV VMI与Rho图像均可用于胰腺放疗剂量计算,Rho图像在碘对比剂区域可减少增强带来的不确定性。
局限性
- 原文未报告具体局限性描述。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Photon-counting computed tomography (PCCT) offers versatile anatomic information because of better energy discrimination and higher spatial resolution than conventional energy-integrating computed tomography (CT). With its rapid applications in diagnostic imaging, its potential within radiation oncology remains largely unexplored. Successful radiation therapy (RT) relies on both high-quality images to delineate tumor volume and accurate physical information of the anatomy for dose calculation. Specific to RT, dose calculation stands as a pivotal component in the process of treatment planning. This proof-of-concept study aimed to assess the accuracy of dose calculation and build the benchmark using PCCT images from a clinically operating NAEOTOM Alpha PCCT scanner from Siemens Healthineers. METHODS AND MATERIALS: A total of 29 patients receiving abdominal contrast-enhanced PCCT scans were included in this study with institutional review board approval. The following 2 sets of reconstructed images were selected for this study: (1) virtual monoenergetic images (VMIs) of 70 keV, which resembles the conventional CT image at 120 kVp, and (2) electron density (Rho) images derived from PCCT, which provides direct physical information for photon dose calculation. In addition to the default Hounsfield look-up table (HLUT) implemented in a treatment planning system, which is suitable for conventional CT images, a specific HLUT converting 70 keV VMI CT numbers to relative electron density was established following the consensus guidelines. Patient's organs were contoured using a deep-learning-based autosegmentation model in syngo.via RT Image Suite from Siemens Healthineers. An elective and a gross tumor volume were simulated in the pancreatic region for treatment planning within this study. All patients were planned with a 9-field intensity modulated RT with simultaneous integral boost regimen 25/33 Gy using an in-house deep-learning-based autoplanning model. Planning guidelines followed the institutional pancreas stereotactic body RT protocol. Eleven dose-volume histogram (DVH) metrics were included for comparison of 3-dose calculation approaches (conventional, 70 keV VMI and Rho). RESULTS: The results revealed minimal dosimetry differences between 70 keV VMI and Rho images, with DVH metric percentage differences predominantly within 1% range (max: -1.60%) and 3-dimensional gamma analysis (1 mm/1%) pass rate >98% for most cases. The patient with the largest differences showed an average DVH metric difference of -1.32% and 3-dimensional gamma pass rate of 94.82%. Overall isodose distribution remained similar between 2 images for each patient. Dose calculation using the default HLUT in Eclipse showed DVH differences within a 1.5% range. CONCLUSIONS: The results demonstrated good agreement in dose calculation between 70 keV VMI and Rho images. For regions with iodine contrast, Rho images provided by PCCT can suppress the contrast enhancement, thereby mitigating unnecessary uncertainties. This benchmark performance shows that PCCT can be safely used for dose calculation with certain advantages, making it a valuable alternative for RT treatment planning.
引用
- Wu X, Wohlfahrt P, Shah J, et al. Accuracy of Photon Dose Calculation on Photon-Counting Computed Tomography-A Comparison Study Based on Virtual Monoenergetic and Electron Density (Rho) Images for Pancreatic Cases. Adv Radiat Oncol. 2025. PMID: 41541756. doi:10.1016/j.adro.2025.101957
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。