基于智能手机的表面引导放疗运动监测系统

A Smartphone-Based Motion Monitoring System for Surface Guided Radiation Therapy.
PubMed 文献 · Adv Radiat Oncol · 2025-12-09

本研究开发并验证了利用iPhone LiDAR传感器的iOS表面引导放疗应用iSGRT,用于患者定位与呼吸监测。静态精度与治疗床位移高度相关(平移r²≥0.995,旋转r²≥0.975),动态精度与模体运动高度一致(r²≥0.963),健康志愿者DIBH时长与SDX肺量计相当(差值0.03秒)。结论认为该低成本系统可行性良好,精度与临床SGRT系统相当。

要点

  • iSGRT基于iPhone LiDAR实现6自由度实时运动追踪,时间分辨率约200-250毫秒(4-5 Hz)。
  • 静态精度:平移r²≥0.995、旋转r²≥0.975,偏差≤0.9 mm、≤0.4°。
  • 健康志愿者DIBH时长与SDX肺量计差值仅0.03秒,动态模体r²≥0.963。

核心结论

  • iOS智能手机表面引导放疗应用iSGRT可提供实时呼吸运动监测,作为可行的替代运动监测系统。
  • iSGRT精度与现有临床SGRT系统相当,同时显著降低成本和复杂度。
  • 该技术有望扩大SGRT可及性,尤其在资源有限环境中。

研究设计与样本

  • 研究为技术开发与验证研究,使用Xcode(Swift和Open3D)开发iSGRT应用。
  • 硬件采用Apple iPhone 15 Pro的LiDAR摄像头,Apple iPad Pro用于远程监测。
  • 静态精度在Varian TrueBeam配PerfectPitch 6DoF治疗床上通过程控床位移与LiDAR位移对比评估。
  • 动态精度使用QUASAR呼吸运动模体,编程正弦及患者来源呼吸波形评估。
  • 健康志愿者在Varian Halcyon孔径内行DIBH,与SDX肺量计系统对比。

主要发现

  • iSGRT与治疗床位移在所有平移轴(r²≥0.995)和旋转轴(r²≥0.975)均呈强相关,偏差≤0.9 mm、≤0.4°。
  • 动态运动评估显示应用与模体真实运动高度一致(r²≥0.963),位移精度对角度的依赖性极小(r²≥0.950)。
  • 健康志愿者DIBH时长在两系统间相当:ΔDIBH = DIBH_SDX - DIBH_iSGRT = 33.34秒 - 33.31秒 = 0.03秒。

关键数值与推荐

  • 时间分辨率约200至250毫秒(4-5 Hz),与临床SGRT系统相当。
  • 静态平移精度r²≥0.995,旋转精度r²≥0.975,偏差≤0.9 mm、≤0.4°。
  • 动态模体运动一致性r²≥0.963,角度依赖性r²≥0.950。
  • 健康志愿者DIBH时长差值0.03秒(SDX 33.34秒 vs iSGRT 33.31秒)。

临床怎么用

  • iSGRT可作为低成本表面引导放疗方案,用于患者定位和呼吸运动监测。
  • 尤其适用于资源有限环境中需要DIBH的左侧乳腺癌放疗等场景。
  • 临床部署前需在更多患者和场景中进一步验证其长期稳定性与工作流程整合。

局限性

  • 研究仅在健康志愿者中进行了DIBH对比,未在患者队列中验证。
  • 动态精度评估使用模体而非真实患者解剖。
  • 尚未报告多中心或低资源环境下的实际部署数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Surface guided radiation therapy (SGRT) improves patient setup and motion monitoring, particularly for deep-inspiratory breath-hold (DIBH) maneuvers in left-sided breast cancer treatment. However, high costs and complexity limit widespread adoption, especially in low-resource settings. In this study, the purpose is to develop and validate a smartphone-based iOS SGRT application (iSGRT) leveraging Light-Detection-and-Ranging (LiDAR) sensors on smartphone-devices for accurate, low-cost surface tracking for radiation therapy. METHODS AND MATERIALS: iSGRT was developed in Xcode using Swift and Open3D, and captures 6-degrees-of-freedom (6DoF) motion for patient positioning and respiratory monitoring. Application was tested using the LiDAR camera on an Apple iPhone 15 Pro, with an Apple iPad Pro for remote monitoring. The system achieved a temporal resolution of ∼200 to 250 ms (4-5 Hz), comparable to clinical SGRT systems. Static accuracy was evaluated by comparing LiDAR-derived displacements with programmed couch movements on a Varian TrueBeam with a PerfectPitch 6DoF couch. Dynamic accuracy was assessed using a QUASAR respiratory motion phantom programmed with sinusoidal and patient-derived breathing waveforms. Motion tracking performance was analyzed using Pearson correlations and Bland-Altman agreement using GraphPad Prism. iSGRT was compared with SDX spirometry system in a healthy volunteer performing DIBH within the bore of a Varian Halcyon. RESULTS: iSGRT demonstrated strong correlations with couch displacements across all translational (r²≥ 0.995) and rotational (r² ≥ 0.975) axes, with minimal biases (≤0.9 mm, ≤0.4°). Dynamic motion evaluation showed high agreement between the application and ground-truth phantom motion (r² ≥ 0.963), with minimal angular dependence on displacement accuracy (r² ≥ 0.950). Breath-hold duration was comparable in the healthy volunteer between systems (ΔDIBH = DIBHSDX - DIBHiSGRT = 33.34seconds - 33.31 seconds = 0.03 seconds). CONCLUSIONS: Feasibility of an iOS smartphone-based SGRT application to provide real-time respiratory motion is demonstrated in this study as a viable alternative motion monitoring system. The iSGRT application's accuracy aligns with existing clinical SGRT systems while significantly reducing cost and complexity. This technology has the potential to expand SGRT accessibility, particularly in resource-limited settings.