头颈再照射患者可执行质子适形FLASH放疗计划研究

Deliverable proton conformal FLASH radiotherapy treatment planning for head and neck re-irradiation patients.
PubMed 文献 · Radiother Oncol · 2025-12-19

本研究为15例头颈再照射患者制定40 GyRBE/5次质子适形FLASH计划,使用商用TPS研究版与临床回旋加速器喷嘴集成硬件,并与VMAT和IMPT进行剂量学比较。所有FLASH计划均满足靶区和危及器官目标,剂量分布与IMPT相似,低剂量区较VMAT减少,CTV平均剂量率在500 nA束流下达95.75±22.78 Gy/s。

要点

  • 15例头颈再照射患者质子适形FLASH计划均满足靶区及OAR目标。
  • FLASH计划剂量分布与临床IMPT相似,低剂量区较VMAT减少。
  • 500 nA下CTV平均剂量率达95.75±22.78 Gy/s。

核心结论

  • 头颈再照射患者可执行质子适形FLASH计划质量与临床IMPT相似且可交付。
  • 机器特异性3D剂量率分布可在TPS中计算并显示。

研究设计与样本

  • 对15例头颈再照射患者制定40 GyRBE/5次靶区质子适形FLASH计划。
  • 使用商用TPS研究版与临床回旋加速器喷嘴集成适形FLASH硬件。
  • TPS配置经验证的超高剂量率束流测量数据生成单束或双束计划。
  • 每束包含单能PBS图、3D打印适形能量调制器、铝距离移位板和黄铜孔径。
  • 用含机器特异性交付定时参数的Python脚本进行蒙特卡罗剂量和剂量率计算。

主要发现

  • 所有FLASH计划均满足肿瘤靶区和OAR计划目标。
  • 适形FLASH计划剂量分布与临床IMPT计划非常相似。
  • 与VMAT相比,IMPT和FLASH计划低剂量区减少。
  • 脊髓最大剂量D0.03 cc:VMAT 8.37±0.94 GyRBE,IMPT 3.19±3.81 GyRBE,FLASH 4.32±3.12 GyRBE。
  • 对侧腮腺平均剂量:VMAT 1.88±0.99 GyRBE,IMPT 0.00±0.01 GyRBE,FLASH 0.00±0.00 GyRBE。

临床怎么用

  • 该硬件配置和束流数据支持下可生成可交付的质子适形FLASH计划。
  • FLASH计划剂量质量与IMPT相似,可作为头颈再照射的潜在技术选项。
  • TPS可计算并显示机器特异性3D剂量率分布供临床评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Clinical translation of ultra-high dose rate (UHDR) delivery to harness potential FLASH effect requires a treatment planning system (TPS) to optimize and calculate dose and dose rate in patients. Proton conformal FLASH treatment aims to deliver pencil beam scanning (PBS) Bragg Peaks to the tumor region with UHDR. In this work, we conducted a treatment planning study for head and neck (H&N) re-irradiation patients using a research-version of a commercial TPS paired with conformal FLASH hardware integrated into a nozzle of a clinical cyclotron-based system. METHODS: Fifteen H&N patients were planned for re-irradiation of 40 GyRBE in 5 fractions to the area of intact tumor. The TPS was configured with validated UHDR beam measurement to generate optimized patient FLASH plans with one or two beams, delivered as single-beam-per-fraction (SBPF). Each beam consists of a deliverable mono-energetic PBS map, a 3D-printable conformal energy modulator design, a selection of aluminum range shifter plates, and a brass aperture. Python scripts with machine-specific delivery timing parameters were used for Monte Carlo dose and dose rate calculations. Clinical VMAT and IMPT plans were also generated for dosimetric comparison. RESULTS: All plans met the tumor target and OAR planning objectives. Conformal FLASH plans showed very similar dose distributions to the clinical IMPT plans. Compared to VMAT plans, both IMPT and FLASH plans have reduced low dose region, maximum cord dose D0.03 cc (8.37 ± 0.94 vs. 3.19 ± 3.81 and 4.32 ± 3.12 GyRBE, respectively), contra-lateral parotid mean dose (1.88 ± 0.99 vs. 0.00 ± 0.01 and 0.00 ± 0.00 GyRBE, respectively) and contra-lateral submandibular gland mean dose (2.49 ± 1.06 vs. 0.14 ± 0.13 and 0.19 ± 0.19 GyRBE, respectively). With 500 nA quasi-continuous nozzle beam current, the mean dose-averaged dose rate in CTVs of these 15 patients achieved 95.75 ± 22.78 Gy/s. CONCLUSIONS: We report the deliverable proton conformal FLASH treatment plans for H&N re-irradiation patients using the innovative hardware configuration and measured beam data in our institution. The FLASH plans have very similar plan qualities to clinical IMPT proton plans and were deliverable with our proton machine. The machine specific 3D dose rate distribution can be calculated and displayed in the TPS.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Zou W, Dong L, Pin A, et al. Deliverable proton conformal FLASH radiotherapy treatment planning for head and neck re-irradiation patients. Radiother Oncol. 2025. PMID: 41423134. doi:10.1016/j.radonc.2025.111349

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。