文献速递放疗物理与技术姑息放疗与转移瘤

二维抗散射栅提升放疗锥形束CT定量精度

A pilot study evaluating 2D antiscatter grid based scatter suppression for quantitative cone beam CT in radiotherapy.
PubMed 文献 · Radiother Oncol · 2025-11-12

该前瞻性预试验纳入31例接受前列腺、盆腔、腹部或头颈部放疗的患者,在同一CBCT-直线加速器上比较标准临床CBCT与采用二维抗散射栅的qCBCT。结果显示qCBCT使CT值绝对误差和伪影幅度分别最多降低56%和53%(p<0.0001),在侧向分离>40 cm患者和下颈部区域改善更显著。

要点

  • qCBCT使CT值绝对误差和伪影幅度分别最多降低56%和53%。
  • 侧向分离>40 cm患者CT值误差降低达62%,伪影幅度降低达60%。
  • 下颈部区域CT值误差降低达72%。

核心结论

  • 基于二维抗散射栅的硬件散射抑制可显著改善CBCT的CT值精度,在侧向分离>40 cm和下颈部区域改善更明显。

研究设计与样本

  • 前瞻性预试验纳入31例接受前列腺、盆腔、腹部或头颈部放疗的患者,在同一CBCT系统上使用相同采集参数分别行标准临床CBCT和采用二维抗散射栅的qCBCT扫描。

主要发现

  • 与临床CBCT相比,qCBCT使CT值绝对误差和伪影幅度分别最多降低56%和53%(p<0.0001)。
  • 侧向分离>40 cm患者中,qCBCT使CT值误差降低达62%、伪影幅度降低达60%(p<0.0001)。
  • 下颈部区域qCBCT的CT值误差降低达72%(p<0.0001)。
  • 相对于qCBCT,迭代重建临床CBCT的CNR高20%(p<0.001),标准临床CBCT的CNR低11%(p<0.05)。

临床怎么用

  • 该硬件散射抑制方法有望提升CBCT用于在线剂量计算、结构分割和疗效评估的定量准确性。
  • 对侧向分离>40 cm或下颈部区域患者,qCBCT的定量精度改善更为显著。

局限性

  • 原文未报告局限性内容。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Cone beam CT (CBCT) images often suffer from poor CT number accuracy, which adversely affects the use of CBCT in online dose calculations, structure segmentation for plan adaptations, and CBCT-based response assessment during radiotherapy. A major cause of inferior CBCT image quality is X-ray scatter. To address this challenge and improve CBCT fidelity, a dedicated two-dimensional (2D) antiscatter grid was developed and its impact on CBCT image quality was evaluated in a prospective trial. This new approach is referred to as quantitative CBCT (qCBCT). METHODS: A total of 31 patients undergoing radiotherapy to the prostate, pelvis, abdomen, or head and neck regions were enrolled and scanned using both a standard-of-care clinical CBCT protocol and the qCBCT, which employed a new 2D antiscatter grid prototype in a C-shaped CBCT-linac. Both scans were performed using identical acquisition parameters on the same CBCT system. For the qCBCT scan, the standard 1D antiscatter grid was exchanged with the 2D antiscatter grid. Clinical CBCT scans were reconstructed using standard and iterative reconstruction algorithms. CT number errors, artifact amplitudes, and contrast-to-noise ratio (CNR) in soft tissues were compared between clinical CBCT and qCBCT. Statistical significance of differences between clinical CBCT and qCBCT image quality metrics was evaluated using a linear mixed model. RESULTS: When compared to clinical CBCT, qCBCT reduced absolute CT number errors and artifact amplitudes by up to 56 % and 53 %, respectively (p < 0.0001). CNR was 20 % higher in clinical CBCT with iterative reconstruction (p < 0.001) and 11 % lower in standard clinical CBCT (p < 0.05), relative to qCBCT. In patients with lateral separation greater than 40 cm, qCBCT yielded up to 62 % lower CT number errors and 60 % lower artifact amplitudes (p < 0.0001). CT number error reduction by qCBCT reached 72 % in the lower neck region (p < 0.0001). CONCLUSION: This pilot trial showed that hardware-based scatter mitigation with a 2D antiscatter grid significantly improved CT number accuracy in CBCT when compared to clinically available CBCT imaging methods. The improvement in quantitative accuracy was more pronounced for patients with lateral separation greater than 40 cm and in the lower neck region.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Bayat F, Sabounchi R, Pyakurel U, et al. A pilot study evaluating 2D antiscatter grid based scatter suppression for quantitative cone beam CT in radiotherapy. Radiother Oncol. 2025. PMID: 41232778. doi:10.1016/j.radonc.2025.111278

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。