患者特异性骨接合板附近辐射剂量不准确:一项离体剂量学研究
该离体剂量学研究使用两块猪胫骨,分别固定CAD/CAM患者特异性2.0 mm重建板或两枚1.0 mm微型板,模拟下颌骨重建后放疗。在软组织等效环境中以2 Gy处方剂量照射,测量七个标准位置的热释光剂量计。结果显示接合板附近实测剂量超过计划剂量高达113%(重建板)和114%(微型板),提示需要优化计划方法。
要点
- 接合板附近实测剂量超过计划剂量高达113%(重建板)和114%(微型板)。
- 接合板上方和骨内出现剂量过高,远离接合板区域剂量过低。
- 术后放疗计划需考虑骨接合材料对剂量分布的影响。
核心结论
- 两种接合板设置中,接合板周围实测剂量均超过计划和处方剂量。
- 研究结果强调需要优化计划方法,并提示骨接合材料在术后放疗计划和骨及周围软组织放射毒性中具有潜在临床转化意义。
研究设计与样本
- 离体剂量学研究使用两块猪胫骨,每块分别固定一个CAD/CAM患者特异性2.0 mm重建板或两枚1.0 mm微型板,模拟下颌骨重建中骨游离瓣的固定。
- 热释光剂量计放置在接合骨内外的七个标准位置,在软组织等效环境中以2 Gy处方剂量照射,测量剂量与计划软件计算剂量比较。
主要发现
- 重建板设置中实测剂量超过计算计划剂量高达113%。
- 微型板设置中实测剂量超过计算计划剂量高达114%。
- 两种设置均观察到接合板上方和骨内剂量过高、远离接合板区域剂量过低的偏差模式。
临床怎么用
- 术后放疗计划时应考虑骨接合材料对局部剂量的影响,并优化计划方法以减小剂量偏差。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND PURPOSE: Osteoradionecrosis (ORN) inside osseous free flaps and native bone is a severe complication in multimodal therapies for head and neck cancer. The tissue proximity of titanium osteosynthesis plates may influence local radiation dose and planning accuracy, yet data from tissue-equivalent, image-guided models are limited. MATERIALS AND METHODS: An ex vivo dosimetric study was performed using two porcine tibiae, each plated with either one computer aided design and manufacturing (CAD/CAM) patient-specific 2.0 mm reconstruction plate or two 1.0 mm mini plates, thus reproducing the fixation of osseous free flaps in primary mandibular reconstruction prior to radiotherapy. Thermoluminescent dosimeters (TLDs) were placed at seven standardized positions inside and outside the plated bones. Irradiation with a prescribed dose of 2 Gy was performed in soft tissue equivalent environment. Measured doses for all TLD positions were acquired and compared to planned doses from the planning software. RESULTS: Local dose enhancements in close proximity to osteosynthesis plates were observed in both setups, with measured doses exceeding calculated planned doses by up to 113% in reconstruction plates and 114% in mini plates. Distinct dose deviation patterns of overdosing above the plates, within the bone and underdosing in plate-distant areas compared to planned and prescribed doses were observed in both setups. CONCLUSION: In both setups, measured doses exceeded planned and prescribed doses around osteosynthesis plates. These findings underline the need for optimized planning approaches and the potential translational clinical relevance of osteosynthesis materials in postoperative radiotherapy planning and radiotoxicity to bone and surrounding soft tissues.
引用
- Fenske J, Neumann O, Nikolaidou E, et al. Radiation dose inaccuracies near patient-specific osteosynthesis plates: an ex vivo dosimetric study. Radiother Oncol. 2025. PMID: 41183681. doi:10.1016/j.radonc.2025.111244
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。