Autostent:半自动设计定制3D打印口腔放疗支架
研究开发MATLAB半自动设计程序Autostent,用于头颈癌患者张口压舌口腔支架的设计。三名用户对四名患者重复设计,半自动法较非自动法平均缩短设计时间23.6分钟(51.2%,p=0.001),支架体积减少4.33 mL(12.9%,p=0.016),并降低用户间体积变异性。
要点
- 半自动化设计使支架设计时间平均缩短51.2%,不受用户、牙科解剖或试验次数影响。
- 半自动化设计使支架体积平均减少12.9%,但多因素分析未达统计学显著。
- 半自动化设计降低了用户间支架体积变异性,平均标准误减少40%。
核心结论
- 半自动设计工作流可显著提高定制3D打印张口压舌支架的设计效率并降低设计变异性。
研究设计与样本
- 基于MATLAB开发半自动设计程序Autostent,与既往非自动计算机辅助设计方法比较。
- 三名用户对四名头颈癌患者重复设计支架三次,比较设计时间和支架体积。
主要发现
- 半自动化使平均设计时间减少23.6分钟(51.2%,p=0.001),不受用户、牙科解剖或试验次数影响。
- 半自动化使平均支架体积减少4.33 mL(12.9%,p=0.016,单因素分析),多因素分析未显著(p=0.40)。
- 半自动化降低用户间支架体积变异性,总体均值标准误减少40%。
临床怎么用
- 半自动设计流程可用于标准化头颈癌放疗口腔支架设计并缩短设计时间。
局限性
- 研究仅纳入四名患者,样本量小。
- 支架体积减少在多因素分析中未达统计学显著。
- 需进一步开发以实现全自动化并优化临床整合。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Oral stents may reduce toxicities during radiation therapy for head and neck cancer (HNC). Customized 3D-printed oral stents offer faster production and achieve comparable patient-reported outcomes to conventionally fabricated stents. However, their design process remains time-consuming, lacks standardization, and relies heavily on skilled technicians. We hypothesized that semi-automating the design process for 3D-printed, mouth-opening, tongue-depressing (MOTD) stents could standardize the design workflow and decrease design time. METHODS: Using oral stent design principles established over decades by oral oncologists, we created a customized computer program (Autostent) using MATLAB to semi-automate the design process of MOTD stents. We subsequently compared Autostent to a previously described method that utilized non-automated computer-aided design. Three users designed stents for four patients with HNC enrolled in a prospective observational study. These patients were selected based on their varying dental anatomies, and each user repeatedly designed an MOTD stent for each patient three times, employing both the non-automated and semi-automated methods. Both methods were compared in terms of design time and stent volume. RESULTS: Semi-automation reduced the average design time by 23.6 min (51.2%, p = 0.001), regardless of user, dental anatomy, or trial number. Additionally, semi-automation decreased the average stent volume by 4.33 mL (12.9%, p = 0.016, univariate analysis). Although this reduction was not statistically significant when considering other experimental variables (p = 0.40, multivariate analysis), semi-automation did lower the variability in stent volume among users (the overall standard error of the mean decreased by 40%). CONCLUSION: Our semi-automated workflow for designing and fabricating customized, 3D-printed MOTD stents significantly improves efficiency and reduces variability in the design. While these results indicate greater consistency compared to manual methods, further development is warranted to achieve full automation and to optimize clinical integration.
引用
- Agrawal A, Tino RB, Zaid M, et al. Autostent: a semi-automated approach to designing customized 3D-printed oral radiation stents for patients with head and neck cancer. Radiat Oncol. 2025. PMID: 41250211. doi:10.1186/s13014-025-02727-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。