文献速递头颈部肿瘤放疗放疗物理与技术

头颈放疗中开放式与封闭式面罩的分次内运动稳定性:OPEN III期试验启示

Intrafraction motion stability of open vs. closed facemasks in head and neck radiotherapy: Insights from the OPEN phase III trial.
PubMed 文献 · Radiother Oncol · 2025-05-21

OPEN III期试验子研究对前56例头颈癌患者进行计划性中期分析,比较封闭面罩与两种开放式面罩固定下的分次内运动。CBCT和SGRT测量显示各面罩类型间分次内运动无临床显著差异,基于CBCT的分次内PTV外扩在各组间相似,开放式面罩分次内稳定性与封闭面罩相当。

要点

  • 开放式与封闭式面罩的分次内运动无临床显著差异。
  • 基于CBCT的分次内PTV外扩为VRT 1.8 mm、LAT 1.7 mm、LNG 1.3 mm。
  • SGRT可检测CBCT未捕捉的瞬态偏差,尤其偏航轴。

核心结论

  • CBCT和SGRT测量的分次内运动均在临床可接受范围内,开放与封闭面罩产生相似PTV外扩。
  • 3点和5点开放式面罩在头颈放疗中实现与封闭面罩相当的分次内稳定性。

研究设计与样本

  • OPEN III期试验子研究,对入组前56例患者进行计划性中期分析作为安全核查。
  • 患者随机分为封闭面罩、3点开放式面罩或5点开放式面罩三组,用CBCT和SGRT评估分次内运动。

主要发现

  • CBCT平均偏差小于0.4 mm和0.2度,SGRT记录所有患者第95百分位偏差为0.4 mm和0.8度。
  • SGRT检测到CBCT未捕捉的瞬态最大偏差,尤其在偏航轴,但差异为瞬态。
  • 贝叶斯分析显示面罩类型或测量方法间分次内运动无临床显著差异。
  • 在记录的小运动范围内,SGRT与CBCT测量的运动无相关性。
  • 各臂间分次内外扩需求无差异。

关键数值与推荐

  • 基于CBCT测量的分次内运动,所有面罩类型计算的外扩为垂直(VRT)1.8 mm、侧向(LAT)1.7 mm、纵向(LNG)1.3 mm。
  • SGRT确认治疗中患者运动保持在这些外扩范围内。

临床怎么用

  • 开放式面罩可用于头颈放疗固定,分次内稳定性与封闭面罩相当。
  • 可参考VRT 1.8 mm、LAT 1.7 mm、LNG 1.3 mm的分次内PTV外扩。
  • SGRT可辅助检测CBCT未捕捉的瞬态和旋转偏差。

局限性

  • 为前56例患者的计划性中期分析,样本量有限。
  • SGRT与CBCT测量的运动在记录的小范围内无相关性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: This OPEN (Optimising Patient Experience in Head and Neck Radiotherapy) phase III trial sub-study, aimed to evaluate intrafraction motion in head and neck (H&N) cancer patients using three different facemask designs. Specifically, we compared intrafraction motion among patients immobilized with a closed facemask or one of two open-face designs, utilizing pre-/post-cone-beam computed tomography (CBCT) and surface-guided radiation therapy (SGRT). METHODS: A pre-planned interim analysis on the first 56 patients enrolled in the OPEN trial was conducted as a safety checkpoint. In the OPEN trial, patients are randomised into three arms: closed facemask, 3-point open facemask, or 5-point open facemask. Intrafraction motion was assessed using both CBCT and SGRT. CBCT provided deviations in translational and rotational dimensions based on bony alignment, while SGRT offered continuous monitoring of surface motion. Intrafraction motion metrics, (i.e. mean, standard deviation, maximum absolute deviation, and the 95th percentile of surface motion) were recorded for each open mask patient using SGRT data to fully quantify motion variation during treatment. The 95th percentile of SGRT deviations was used for direct comparison with CBCT motion data. Bayesian analysis was conducted to determine the equivalence of motion across mask types and measurement techniques. Margins to account for intrafraction motion were calculated across mask types using Van Herk's formulism. RESULTS: Mean CBCT deviations were less than 0.4 mm and 0.2 degrees, while SGRT recorded 95th percentile deviations of 0.4 mm and 0.8 degrees over all patients. SGRT detected transient maximum deviations not captured by CBCT, particularly in the yaw axis. However, these differences were transient. Bayesian analysis showed no clinically significant differences in intrafraction motion between mask types or measurement methods. No correlation was found between SGRT and CBCT measured motion within the small range of motion recorded. No difference in intrafraction margin requirements were found between arms. Based on CBCT-measured intrafraction motion, margins of 1.8 mm, 1.7 mm, and 1.3 mm were calculated for the vertical (VRT), lateral (LAT), and longitudinal (LNG) directions, respectively, to account for intrafraction motion for all mask types, with SGRT confirming that patient motion during treatment remained within these margins. CONCLUSIONS: Intrafraction motion, as measured by both CBCT and SGRT, remains within clinically acceptable limits and yields similar PTV margins across both open and closed mask types. Intrafraction PTV margins were found to be comparable across all mask types. The use of SGRT allowed for the detection of transient deviations and rotational differences that were not detected using CBCT alone. Overall, these findings confirm that both 3-point and 5-point open-face masks achieve intrafraction stability comparable to traditional closed facemasks in head and neck radiotherapy.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Malone C, Ryan S, Nicholson J, et al. Intrafraction motion stability of open vs. closed facemasks in head and neck radiotherapy: Insights from the OPEN phase III trial. Radiother Oncol. 2025. PMID: 40409364. doi:10.1016/j.radonc.2025.110941

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。