3D打印补偿器用于儿童浅表肿瘤质子笔束扫描
本研究基于六例头颈部肉瘤患儿的回顾性数据,利用治疗计划系统模拟比较射程移位器与3D打印补偿器在质子笔束扫描中的剂量学差异。结果显示补偿器可减少低剂量照射区、改善适形度并降低危及器官暴露,侧向束斑尺寸减小约57%,半影减小41-47%。
要点
- 3D打印补偿器使侧向束斑尺寸减小约57%,不同深度半影减小41-47%。
- 补偿器可减少低剂量照射区域、改善适形度并降低危及器官暴露。
- 补偿器均匀性变化小于3.5%,满足计划稳健性评估标准。
核心结论
- 与射程移位器置于喷嘴处相比,将3D打印补偿器置于患者附近治疗浅表肿瘤可获得更适形的剂量分布。
研究设计与样本
- 基于治疗计划系统模拟,回顾性选取六例头颈部肉瘤患儿数据,其中三例在治疗阶段已使用3D打印补偿器。
- 比较使用射程移位器与补偿器的治疗计划,评估靶区约束(D98%>95%、D2%<107%)及危及器官约束(D2%、Dmax、DMean)。
- 呈现补偿器制作全流程,包括熔融长丝制造3D打印、材料水等效比测量及均匀性、打印质量和CT值评估。
主要发现
- 补偿器减小了低剂量照射区域、改善了适形度并降低了关键器官暴露。
- 补偿器使立方体靶区内不同深度的侧向束斑尺寸减小约57%,半影减小41-47%。
- 补偿器均匀性变化小于3.5%,满足计划稳健性评估标准。
关键数值与推荐
- 靶区约束:D98%>95%、D2%<107%。
- 危及器官(脑干、脊髓、视觉器官、视交叉、耳蜗)约束:D2%、Dmax、DMean。
临床怎么用
- 对于浅表病变患儿,可考虑使用患者特异性3D打印补偿器以减小空气间隙和束流散射。
- 使用补偿器前需完成材料水等效比测量及均匀性、打印质量、CT值评估。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: In modern proton radiotherapy facilities with pencil beam scanning technology, the lowest energy of a proton beam typically ranges between 60 and 100 MeV, corresponding to a proton range in water of 3.1-7.5 cm. The irradiation of superficial lesions usually requires the application of a range shifter (RS) to further reduce the proton range. A certain distance from the patient to the RS increases the spot size, causing worse plan conformity. As an alternative solution, a patient-specific 3D-printed proton beam compensator (BC) can be applied to reduce the air gap and beam scattering. MATERIALS AND METHODS: This study is based on treatment planning system simulations using retrospectively selected data from six pediatric patients with diagnosed sarcomas located in the head and neck area. For three of these patients, 3D-printed compensators were utilized during the treatment phase, prior to the retrospective analysis. Treatment plans for children with shallow lesions treated using RSs and BCs were compared. Planning target volume constraints (D98% >95%, D2%< 107%) and organs-at-risk (brainstem, spinal cord, visual organs, chiasm, cochlea) constraints (D2%, Dmax and DMean) were applied. The entire process of using a BCs in the treatment of pediatric superficial tumors is presented, including 3D printing procedure (via fused filament fabrication method), dosimetric verification of the material (Water Equivalent Ratio measurements) and assessment of its homogeneity, print quality and Hounsfield Unit specification. Beam parameters analysis including spot sizes and penumbras, were performed. Treatment plans were compared in terms of plan conformity and sparing of critical organs. RESULTS: The application of BCs reduced the low-dose irradiation areas, improved conformity and reduced critical organs exposure. BCs decreased the lateral spot size by approximately 57% and the penumbras by 41-47% at different depths in the cube target. The variation in BC homogeneity was less than 3.5%, meeting the criteria for plan robustness evaluation. CONCLUSIONS: Compared with RS placement at the nozzle, the placement of 3D-printed BCs in the near vicinity of the patient for the treatment of superficial tumors led to a more conformal dose distribution.
引用
- Wochnik A, Kajdrowicz T, Foltyńska G, et al. Application of 3D-printed compensators for proton pencil beam scanning of shallowly localized pediatric tumors. Radiat Oncol. 2025. PMID: 40301908. doi:10.1186/s13014-025-02646-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。