常规直线加速器在线自适应五分割消融放疗治疗前列腺癌

Online Adaptive Five-Fraction Ablative Radiotherapy for Prostate Cancer Using a Conventional Linear Accelerator.
PubMed 文献 · Pract Radiat Oncol · 2025-02-12

该研究评估了在西班牙巴塞罗那Quirónsalud医院使用常规直线加速器对前列腺癌患者进行在线自适应五分割立体定向体部放疗的剂量学结果。对前20例患者的100次在线自适应计划进行分析,并与常规图像引导放疗计划比较。在线自适应方法在所有治疗分次中保证了最佳靶区覆盖,而模拟的常规图像引导放疗计划在70%的分次中未能达到前列腺覆盖V40 Gy ≥ 95%的要求。

要点

  • 在线自适应计划在所有分次中均保证了前列腺靶区的最佳覆盖。
  • 模拟的常规图像引导放疗计划在70%分次中未达到V40 Gy ≥ 95%覆盖标准。
  • 计划适应平均耗时20.1 ± 6.1分钟,总治疗时间平均26.0 ± 6.3分钟。

核心结论

  • 使用常规直线加速器的在线自适应方案治疗前列腺癌临床可行,并符合PACE B试验设定的验收标准。

研究设计与样本

  • 研究在西班牙巴塞罗那Quirónsalud医院开展,对前20例接受五分割前列腺立体定向体部放疗的患者进行在线自适应计划剂量学评估。
  • 每次治疗分次生成基于锥形束CT的自适应「当日计划」,由单弧容积旋转调强放疗构成,共分析100个计划。
  • 每个自适应计划与对应的常规图像引导放疗模拟计划进行比较,并分析自适应治疗时间。

主要发现

  • 在线自适应方法在所有患者的全部分次中保证了最佳靶区覆盖。
  • 模拟的常规图像引导放疗计划在70%的分次中未达到前列腺覆盖V40 Gy ≥ 95%的要求。
  • 自适应计划与常规图像引导放疗计划在危及器官剂量上平均差异较小;部分模拟计划对直肠和膀胱的保护与前列腺覆盖受损相关。
  • 患者特异性质量保证显示在线计算计划与独立剂量核查在全部100个计划中高度一致。
  • 计划适应平均耗时20.1 ± 6.1分钟,包括适应和治疗实施的总治疗时间平均26.0 ± 6.3分钟。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To assess the dosimetric results of an online adaptive stereotactic body radiation therapy (SBRT) program to treat patients with prostate cancer using a conventional linear accelerator. METHODS AND MATERIALS: Prostate SBRT with 5 fractions is performed at the Hospital Quirónsalud Barcelona using an online adaptive method previously described (Pract Radiat Oncol. 2022 Mar-Apr;12(2):e144-e152). A CBCT-based adaptive "plan-of-the-day" (POD) is generated for each treatment fraction, which consists of a single volumetric modulated arc therapy. A dosimetric evaluation of the PODs was performed for the first 20 patients and included analysis of the target (prostate), organs-at-risks (OARs), and patient-specific quality assurance (PSQA). Each of the PODs was also compared with the corresponding conventional image guided radiation therapy (IGRT) method. Finally, the adaptive treatment timing is analyzed for the 100 PODs in this study. RESULTS: The online adaptive treatment method ensured optimal target coverage in all treatment fractions for all patients. However, the simulated IGRT plans did not result in adequate prostate coverage (V40 Gy ≥ 95%) in 70% of fractions. Small average differences between PODs and IGRT plans were found in the OARs. The dose sparing in the rectum and bladder provided by some simulated IGRT plans, was associated with a compromised prostate coverage. The PSQA resulted in an excellent agreement between the online-calculated plans and the independent dose checks performed for all 100 PODs. The average duration of the plan adaptation was 20.1 ± 6.1 minutes and the average overall session time including adaptation and treatment delivery was 26.0 ± 6.3 minutes. CONCLUSIONS: The online adaptive program using a conventional linac to treat prostate cancer described in this study is clinically feasible and in adherence with the acceptance criteria set by the PACE B trial.