基于实时三维光学表面成像系统的乳腺癌放疗中自主深吸气屏气与自由呼吸的可重复性与稳定性

Reproducibility and stability of voluntary deep inspiration breath hold and free breath in breast radiotherapy based on real-time 3-dimensional optical surface imaging system.
PubMed 文献 · Radiat Oncol · 2024-11-11

本研究回顾性纳入17例采用野中野技术行深吸气屏气(DIBH)放疗的患者(510次屏气治疗)和20例采用容积旋转调强放疗(VMAT)行自由呼吸(FB)的患者(600次自由呼吸治疗),基于光学表面成像系统(OSIS)评估分次间可重复性与分次内稳定性。结果显示DIBH与FB在分次间和分次内均表现良好;当患者呼吸运动较大时,FB技术更可能出现靶区覆盖不足,而DIBH技术更可能导致危及器官(肺、心脏和对侧乳腺)剂量增加。

要点

  • OSIS与CBCT在DIBH和FB治疗中各方向一致性均小于0.30 cm。
  • DIBH分次内呼吸运动在侧向、纵向、垂直方向分别为-0.06±0.07 cm、0.12±0.15 cm、0.12±0.12 cm。
  • FB分次内呼吸运动在侧向、纵向、垂直方向分别为-0.02±0.12 cm、0.08±0.18 cm、0.14±0.20 cm。

核心结论

  • 基于OSIS,DIBH与FB均能在分次间提供良好的可重复性并在分次内保持稳定性。
  • 当患者呼吸运动较大时,FB技术更可能出现靶区覆盖不足,而DIBH技术更可能导致危及器官(肺、心脏和对侧乳腺)剂量增加。

研究设计与样本

  • 回顾性研究,纳入17例采用野中野(FiF)技术行DIBH放疗的患者(510次屏气治疗)和20例采用VMAT技术行FB放疗的患者(600次自由呼吸治疗)。
  • 所有患者均在CBCT和OSIS引导下摆位,并在整个治疗过程中接受OSIS监测。
  • 从OSIS治疗报告中提取所有治疗分次的三个方向共8种摆位变异,并手动引入治疗计划,共生成296个扰动计划,重新计算并分析靶区和危及器官的剂量体积直方图(DVH)。

主要发现

  • OSIS与CBCT在DIBH和FB治疗中各方向一致性均小于0.30 cm。
  • DIBH分次内呼吸运动在侧向、纵向、垂直方向分别为-0.06±0.07 cm、0.12±0.15 cm、0.12±0.12 cm;FB分别为-0.02±0.12 cm、0.08±0.18 cm、0.14±0.20 cm。
  • 对靶区而言,DIBH计划对摆位误差更敏感;CTV的D95平均偏差在DIBH为39.78 Gy-40.17 Gy,在FB为38.46 Gy-40.52 Gy。
  • 对危及器官而言,FB计划的心脏V10、V20和Dmean,同侧肺V5、V20和Dmean,以及乳腺Dmean的平均偏差均低于DIBH计划。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The aim of this study was to evaluate the inter-fraction reproducibility and intra-fraction stability of breast radiotherapy using voluntary deep-inspiration breath hold (DIBH) and free breathing (FB) based on an optical surface imaging system (OSIS). METHODS: Seventeen patients (510 breath-hold sessions) treated using a field-in-field (FiF) technique and twenty patients (600 breath-free sessions) treated with a volume-modulated arc therapy (VMAT) technique were included in this retrospective study. All the patients were positioned with the guidance of CBCT and OSIS, and also monitored with OSIS throughout the whole treatment session. Eight setup variations in three directions were extracted from the treatment reports of OSIS for all sessions and were subsequently manually introduced to treatment plans, resulting in a total of 296 perturbed plans. All perturbed plans were recalculated, and the dose volume histograms (DVH) for the target and organs at risk (OAR) were analyzed. RESULTS: The OSIS and CBCT for both DIBH and FB treatments showed a good agreement of less than 0.30 cm in each direction. The intra-fraction respiratory motion data during DIBH were -0.06 ± 0.07 cm, 0.12 ± 0.15 cm, and 0.12 ± 0.12 cm in the lateral, longitudinal, and vertical directions, respectively; for FB, the respiratory motion data were -0.02 ± 0.12 cm, 0.08 ± 0.18 cm, and 0.14 ± 0.20 cm, respectively. For the target, DIBH plans were more sensitive to setup errors; the mean deviations in D95 for CTV were 39.78 Gy-40.17 Gy for DIBH and 38.46 Gy-40.52 Gy for FB, respectively. For the OARs, the mean deviations of V10, V20, and Dmean to the heart; V5, V20, and Dmean to the ipsilateral lung; and Dmean to the breast were lower for the FB plan compared with the DIBH plan. CONCLUSION: Based on OSIS, our results indicate that both DIBH and FB can provide good reproducibility in the inter-fractions and stability in the intra-fractions. When the patient respiratory motion is large, the FB technology has greater possibility for the undercoverage of the target volume, while DIBH technology is more likely to result in increases in dose to OARs (the lung, heart, and contralateral breast).