光子计数计算机断层扫描在放射治疗定量成像中的评估
在头颈部模体上比较第一代临床光子计数CT(PCCT)与单能/双能CT(SECT/DECT)对材料定量表征的能力,评估线性衰减系数、相对电子密度、有效原子序数和质子阻止本领比。PCCT在多数材料上分布更窄,但水样材料的有效原子序数和阻止本领比与双能CT相当,钛的电子密度和阻止本领比估计劣于双能CT。作者认为结果有前景,但对临床实践的意义仍需证明。
要点
- PCCT在相对电子密度上所有材料均显示比两种双能CT方法更窄的分布。
- 水样材料的有效原子序数和阻止本领比在PCCT与双能CT间相当。
- 钛的电子密度和阻止本领比估计PCCT劣于双能CT。
核心结论
- 在类人几何模体中,PCCT用于放射治疗的组织表征显示出有前景的结果,但对临床实践改进的意义仍需证明。
研究设计与样本
- 使用CIRS 731-HN头颈部模体,在NAEOTOM Alpha PCCT和SOMATOM Definition AS+上分别以单能CT和双能CT技术扫描,两台设备均由Siemens Healthineers制造。
主要发现
- 对线性衰减系数、相对电子密度/有效原子序数和质子阻止本领比,PCCT相比两种双能CT方法有分布更窄的趋势,但仅在相对电子密度上所有材料均观察到。
- 均方根偏差总体显示PCCT更优,但水样材料的有效原子序数和阻止本领比在各模态间相当,钛的相对电子密度和阻止本领比估计PCCT劣于双能CT。
- 在更复杂的示例射线路径上PCCT数据与理论值的偏差最小,而更标准的投影显示各模态结果相似。
局限性
- 作者指出,PCCT改进对临床实践的意义仍需证明。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To test a first-generation clinical photon-counting computed tomography (PCCT) scanner's capabilities to characterize materials in an anthropomorphic head phantom for radiation therapy purposes. METHODS AND MATERIALS: A CIRS 731-HN head-and-neck phantom (CIRS/SunNuclear) was scanned on a NAEOTOM Alpha PCCT and a SOMATOM Definition AS+ with single-energy and dual-energy CT techniques (SECT and DECT, respectively), both scanners manufactured by Siemens (Siemens Healthineers). A method was developed to derive relative electron density (RED) and effective atomic number (EAN) from linear attenuation coefficients (LACs) of virtual mono-energetic images and applied for the PCCT and DECT data. For DECT, Siemens' syngo.via "Rho/Z"-algorithm was also used. Proton stopping-power ratios (SPRs) were calculated based on RED/EAN with the Bethe equation. For SECT, a stoichiometric calibration to SPR was used. Nine materials in the phantom were segmented, excluding border pixels. Distributions and root-mean-square deviations within the material regions were evaluated for LAC, RED/EAN, and SPR, respectively. Two example ray projections were also examined for LAC, SPR, and water-equivalent thickness, for illustrations of a more treatment-like scenario. RESULTS: There was a tendency toward narrower distributions for PCCT compared with both DECT methods for the investigated quantities, observed across all materials for RED only. Likewise the scored root-mean-square deviations showed overall superiority for PCCT with a few exceptions: for water-like materials, EAN and SPR were comparable between the modalities; for titanium, the RED and SPR estimates were inferior for PCCT. The PCCT data gave the smallest deviations from theoretic along the more complex example ray profile, whereas the more standard projection showed similar results between the modalities. CONCLUSIONS: This study shows promising results for tissue characterization in a human-like geometry for radiation therapy purposes using PCCT. The significance of improvements for clinical practice remains to be demonstrated.
引用
- Zimmerman J, Poludniowski G. Assessment of Photon-Counting Computed Tomography for Quantitative Imaging in Radiation Therapy. Int J Radiat Oncol Biol Phys. 2024. PMID: 39549761. doi:10.1016/j.ijrobp.2024.11.069