利用Halcyon特定复杂度指标通过机器学习和深度学习预测乳腺癌IMRT放疗计划的患者特异性质量保证
研究使用56例乳腺癌患者的318个射束,以7个复杂度指标为输入,建立机器学习(ML)和深度学习(DL)模型预测DreamDose质量保证(QA)符合性;MCS、BI、机架角和准直器角与QA符合性不相关,最终用SAS10、BA和BM训练;DL模型AUC为0.95,特异度和灵敏度分别为0.98和0.97,优于ML模型(AUC 0.75,特异度0.88,灵敏度0.86)。
要点
- MCS、BI、机架角和准直器角与QA符合性不相关。
- DL模型预测QA的AUC为0.95,特异度0.98,灵敏度0.97。
- ML模型AUC为0.75,特异度0.88,灵敏度0.86。
核心结论
- DL模型在预测质量保证结果方面表现出高度准确性。
- 在线预测QA平台在加速器占用和工作时间方面节省显著时间。
研究设计与样本
- 研究使用56例乳腺癌患者的318个射束。
- 以MCS、SAS10、BA、BI、BM、机架角和准直器角共7个复杂度指标作为AI模型输入。
- 使用tensorflow建立机器学习(ML)和深度学习(DL)模型预测DreamDose QA符合性。
主要发现
- MCS、BI、机架角和准直器角与QA符合性不相关。
- ML和DL模型最终使用SAS10、BA和BM复杂度指标进行训练。
- ROC分析确定最佳预测概率阈值以提高特异度和灵敏度。
- ML模型AUC为0.75,特异度和灵敏度分别为0.88和0.86,未显示满意性能。
- 优化后的DL模型AUC为0.95,特异度和灵敏度分别为0.98和0.97,表现更好。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: New radiotherapy machines such as Halcyon are capable of delivering dose-rate of 600 monitor-units per minute, allowing large numbers of patients treated per day. However, patient-specific quality assurance (QA) is still required, which dramatically decrease machine availability. Innovative artificial intelligence (AI) algorithms could predict QA result based on complexity metrics. However, no AI solution exists for Halcyon machines and the complexity metrics to be used have not been definitively determined. The aim of this study was to develop an AI solution capable of firstly determining the complexity indices to be obtained and secondly predicting patient-specific QA in a routine clinical setting. METHODS: Three hundred and eighteen beams from 56 patients with breast cancer were used. The seven complexity indices named Modulation-Complexity-Score (MCS), Small-Aperture-Score (SAS10), Beam-Area (BA), Beam-Irregularity (BI), Beam-Modulation (BM), Gantry and Collimator angles were used as input to the AI model. Machine learning (ML) and deep learning (DL) models using tensorflow were set up to predict DreamDose QA conformance. RESULTS: MCS, BI, gantry and collimator angle are not correlated with QA compliance. Therefore, ML and DL models were trained using SAS10, BA and BM complexity indices. ROC analyses enabled to find best predicted probability threshold to increase specificity and sensitivity. ML models did not show satisfactory performance with an area under-the-curve (AUC) of 0.75 and specificity and sensitivity of 0.88 and 0.86. However, optimised DL model showed better performance with an AUC of 0.95 and specificity and sensitivity of 0.98 and 0.97. CONCLUSION: The DL model demonstrated a high degree of accuracy in its predictions of the quality assurance (QA) results. Our online predictive QA-platform offers significant time savings in terms of accelerator occupancy and working time.
引用
- Boutry C, Moreau NN, Jaudet C, et al. Machine learning and deep learning prediction of patient specific quality assurance in breast IMRT radiotherapy plans using Halcyon specific complexity indices. Radiother Oncol. 2024. PMID: 39159677. doi:10.1016/j.radonc.2024.110483
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。