文献速递乳腺癌放疗放疗物理与技术

深吸气屏气使锥形束成像上的手术腔配准更便利,用于部分乳腺照射

Deep Inspiration breath Hold facilitates surgical cavity registration on cone beam imaging for Partial breast irradiation.
PubMed 文献 · Radiother Oncol · 2024-08-08

本研究评估深吸气屏气(DIBH)联合皮肤引导放疗能否改善锥形束CT(CBCT)图像质量、减少运动伪影并提高手术腔识别与配准一致性。DIBH下腔可视化评分与图像质量均显著改善(p<0.001),手术腔识别率由56%升至76%(p<0.001),配准移位观察者间变异显著降低(p<0.001)。

要点

  • DIBH下腔可视化评分与图像质量Likert评分均较非DIBH显著改善(p<0.001)。
  • 手术腔识别率DIBH组76% vs 非DIBH组56%(p<0.001)。
  • DIBH显著降低配准移位校正的观察者间变异(p<0.001)。

核心结论

  • 利用DIBH进行运动控制可改善图像质量与手术腔识别,进而降低配准变异,这对加速部分乳腺照射(APBI)的准确性具有重要意义。

研究设计与样本

  • 研究纳入CT模拟上手术腔显示清晰的左、右侧乳腺癌患者,采集每日CBCT图像;仅左侧CBCT在DIBH下采集。
  • 由受过培训且有经验的放疗治疗师使用腔可视化评分(CVS)、图像质量Likert评分进行评估并执行配准移位,图像匿名且治疗师对是否使用DIBH设盲。
  • 共21例患者、每例15幅CBCT,由6名放疗治疗师评估,生成4015个评估点。

主要发现

  • DIBH下CVS与图像质量均观察到统计学显著改善(p<0.001)。
  • 手术腔识别率DIBH组为76%,非DIBH组为56%(p<0.001)。
  • DIBH显著降低配准移位校正的观察者间变异(p<0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: The quality of the Cone Beam Computed Tomography (CBCT) images used for patient set-up is essential to avoid geographical miss when narrower margins or shorter fractionation are used for example in Accelerated Partial Breast Irradiation (APBI). This study evaluates deep inspiration breath hold (DIBH) with skin guided radiotherapy as a tool for image improvement reducing motion artifacts. MATERIALS AND METHODS: Daily CBCT images of left and right breast cancer patients with well-defined surgical cavity on CT simulation were used for this study. Only left sided CBCT were acquired with DIBH. Trained and experienced radiation therapists were asked to evaluate the image quality using a cavity visualization score (CVS), an image quality Likert score, and to perform registration shifts. Images were anonymized and therapists were blinded to the use of DIBH. RESULTS: Images from 21 patients, with 15 CBCT each, were evaluated by 6 radiation therapists, generating 4,015 evaluation points. Statistically significant improvements were observed in CVS and image quality (p < 0.001) with DIBH. Also, the rate of surgical cavity identification increased to 76 % with DIBH compared to 56 % without (p < 0.001). DIBH significantly reduced the inter-observer variability in registration shift corrections (p < 0.001) CONCLUSION: Utilizing DIBH for motion control improves both the image quality and the surgical cavity identification. This results in a decrease in registration variability, which is important for APBI accuracy.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Ali T, Vellengara M, Albalushi FY, et al. Deep Inspiration breath Hold facilitates surgical cavity registration on cone beam imaging for Partial breast irradiation. Radiother Oncol. 2024. PMID: 39127406. doi:10.1016/j.radonc.2024.110471

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。