深度学习临床靶区自动勾画在磁共振引导直肠癌在线自适应放疗中的评估
本研究提出一套三步评估框架(定量性能与观察者间变异比较、专家观察与修正、对靶区覆盖率的影响评估),并应用于本院训练的nnU-Net模型,使用44例在本机构接受磁共振引导在线自适应放疗的直肠癌患者数据。框架确定模型经专家修正后性能与观察者间变异相当,虽引入偏倚但对临床实践无相关影响,且在不降低靶区覆盖率的前提下显著节省时间。作者结论为该模型符合临床使用条件。
要点
- 三步评估框架用于深度学习自动勾画模型的临床引入评估。
- 模型经专家修正后性能与观察者间变异相当。
- 模型引入偏倚但对临床实践无相关影响,且节省时间不降低靶区覆盖率。
核心结论
- 基于评估结果,作者结论为该模型符合临床使用条件。
研究设计与样本
- 评估框架包含三步:模型性能定量评估并与观察者间变异比较、专家观察与修正、对预期体积靶区覆盖率的影响评估,各步骤在独立数据集上进行。
- 框架应用于本院训练的nnU-Net模型,使用44例在本机构接受治疗的直肠癌患者数据。
主要发现
- 框架确定模型经专家修正后性能与观察者间变异相当。
- 模型引入偏倚,但该偏倚对临床实践无相关影响。
- 在不降低由体积靶区覆盖率判定的质量的前提下,获得显著的时间节省。
临床怎么用
- 作者结论为该模型符合临床使用条件。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Segmentation of clinical target volumes (CTV) on medical images can be time-consuming and is prone to interobserver variation (IOV). This is a problem for online adaptive radiation therapy, where CTV segmentation must be performed every treatment fraction, leading to longer treatment times and logistic challenges. Deep learning (DL)-based auto-contouring has the potential to speed up CTV contouring, but its current clinical use is limited. One reason for this is that it can be time-consuming to verify the accuracy of CTV contours produced using auto-contouring, and there is a risk of bias being introduced. To be accepted by clinicians, auto-contouring must be trustworthy. Therefore, there is a need for a comprehensive commissioning framework when introducing DL-based auto-contouring in clinical practice. We present such a framework and apply it to an in-house developed DL model for auto-contouring of the CTV in rectal cancer patients treated with MRI-guided online adaptive radiation therapy. METHODS AND MATERIALS: The framework for evaluating DL-based auto-contouring consisted of 3 steps: (1) Quantitative evaluation of the model's performance and comparison with IOV; (2) Expert observations and corrections; and (3) Evaluation of the impact on expected volumetric target coverage. These steps were performed on independent data sets. The framework was applied to an in-house trained nnU-Net model, using the data of 44 rectal cancer patients treated at our institution. RESULTS: The framework established that the model's performance after expert corrections was comparable to IOV, and although the model introduced a bias, this had no relevant impact on clinical practice. Additionally, we found a substantial time gain without reducing quality as determined by volumetric target coverage. CONCLUSIONS: Our framework provides a comprehensive evaluation of the performance and clinical usability of target auto-contouring models. Based on the results, we conclude that the model is eligible for clinical use.
引用
- Ferreira Silvério N, van den Wollenberg W, Betgen A, et al. Evaluation of Deep Learning Clinical Target Volumes Auto-Contouring for Magnetic Resonance Imaging-Guided Online Adaptive Treatment of Rectal Cancer. Adv Radiat Oncol. 2024. PMID: 38706833. doi:10.1016/j.adro.2024.101483
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。