利用自动化追踪工具改善放射治疗直线加速器质量保证程序的管理与依从性
该研究在一家拥有7台直线加速器的大型学术中心实施自研质量保证(QA)仪表板,分析2016至2022年间76,066条记录,评估其对QA依从性和测量变异的影响。QA依从性从2016年的75%提升至2019年实施仪表板后的>99%,影像QA变异系数约降低一半,剂量测量方差无实质变化。
要点
- 实施自动化QA仪表板后,整体QA依从性从2016年75%升至2019年>99%。
- 影像QA的变异系数在实施仪表板后约降低一半。
- 记录的剂量测量方差未见实质变化。
核心结论
- 实施QA仪表板使QA依从性显著提高,所有影像模态的QA测量方差均观察到降低,表明自动化追踪QA工具对改善依从性和准确性有重要影响。
研究设计与样本
- 研究在一家拥有7台直线加速器的大型学术中心开展,2019年实施自研QA仪表板,收集2016至2022年月度QA数据以分析实施前后的依从性。
- 共分析76,066条记录,其中73,187条QA测量按时完成;依从性定义为按时完成测试的百分比。
主要发现
- 整体QA依从性从2016年的75%上升至2019年实施QA仪表板后的>99%。
- 依从性改善主要见于较近期实施的测试,例如影像QA和物理师每日QA审查的正式记录。
- 实施QA仪表板后,影像QA的变异系数约降低一半(约2倍)。
- 对于记录的剂量测量,未观察到方差有实质变化。
关键数值与推荐
- 依从性定义为按时完成测试的百分比。
- 共分析76,066条记录,其中73,187条QA测量按时完成。
- 整体QA依从性从2016年75%升至2019年>99%。
- 影像QA变异系数在实施QA仪表板后约降低一半;剂量测量方差无实质变化。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Although quality assurance (QA) is crucial in radiation oncology departments, substantial efforts are required to monitor and ensure compliance to high standards. This work aims to analyze the impact of implementing a centralized and automated tracking dashboard on compliance and variance observed in radiation therapy QA results for linear accelerators. METHODS AND MATERIALS: The study was performed in a large academic center with 7 linear accelerators. An in-house QA Dashboard was implemented in 2019 with design specifications including automated monitoring and visualization of QA progress, ease of use and accessibility, ease of integration, adaptability to new technologies, and facilitation of automated reminders. Monthly QA data were collected between 2016 and 2022 to analyze compliance pre- and post-dashboard implementation. Compliance was characterized as the percentage of tests completed on time. In addition, variance trends for linear accelerator dosimetry measurements and imaging were analyzed over 7 years. RESULTS: In total, 76,066 records were analyzed. Of these records, 73,187 QA measurements were completed on time. Overall QA compliance increased from 75% in 2016 to >99% in 2019 after successful implementation of the QA Dashboard. The main improvement was observed for tests that were implemented more recently, eg, imaging QA and formal recording of daily QA review by physicists. The coefficient of variation was reduced by approximately a factor of 2 for imaging QA after the implementation of the QA Dashboard. For recorded dosimetry measurements, no substantial change in variance was observed. CONCLUSIONS: Implementation of the QA Dashboard resulted in a distinct increase in QA compliance. Reduction in the variance in QA measurements was observed for all imaging modalities. These findings demonstrate high impact of automated tracking QA tools on improved compliance and accuracy of QA.
引用
- Lyatskaya Y, Killoran J, Kukluk J, et al. Improving Management and Compliance of Radiation Therapy Linear Accelerator Quality Assurance Program With Automated Tracking Tools. Adv Radiat Oncol. 2024. PMID: 38550367. doi:10.1016/j.adro.2024.101469
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。