DIBH乳腺癌放疗中表面引导自动VMAT门控的临床实用性与患者表现
该回顾性研究评估了33例左侧乳腺癌患者采用DIBH与VMAT自动表面引导门控的性能,共分析525个分次。中位DIBH持续51秒,每分次中位需4次屏气完成剂量投照,分次内稳定性与重复性约1.0毫米。模拟显示门控窗容差可从±3.0毫米降至±2.0毫米而不影响束流状态。
要点
- 中位DIBH持续时间51秒,每分次中位需4次屏气完成VMAT剂量投照
- 分次内DIBH稳定性与分次间重复性在各方向约1.0毫米
- 有无音频指导的患者间未发现大的差异
核心结论
- 无论是否使用音频指导,本研究表明DIBH联合VMAT的自动表面引导门控具有高效率,患者DIBH表现远超既往经验与文献预期,且门控窗容差可降低。
研究设计与样本
- 回顾性分析DIBH联合VMAT开展第一年内治疗的左侧乳腺癌患者,共33例患者、525个分次。
- 自动表面引导门控在感兴趣表面区域移出门控容差(±3毫米、±3°)时自动关闭束流,移回时自动开启。
- 指导患者尽可能舒适地屏气,根据患者偏好决定治疗投照时是否给予音频指导。
- 采集乳腺/胸壁表面相对参考表面的三个正交方向实时位置变化,并模拟±2.5毫米、±2.0毫米、±1.5毫米的较小容差以评估最佳门控窗阈值。
主要发现
- 中位DIBH持续时间为51秒(范围30-121秒),完成VMAT剂量投照每分次中位需4次DIBH。
- 分次内DIBH稳定性和分次间DIBH重复性在各方向约1.0毫米。
- 偏好音频指导的患者(n=21)与无音频指导的患者(n=12)之间未发现大的差异。
- 模拟显示门控窗容差可从±3.0毫米降至±2.0毫米而不影响束流状态。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND PURPOSE: To evaluate the performance of automated surface-guided gating for left-sided breast cancer with DIBH and VMAT. MATERIALS AND METHODS: Patients treated in the first year after introduction of DIBH with VMAT were retrospectively considered for analysis. With automated surface-guided gating the beam automatically switches on/off, if the surface region of interest moved in/out the gating tolerance (±3 mm, ±3°). Patients were coached to hold their breath as long as comfortably possible. Depending on the patient's preference, patients received audio instructions during treatment delivery. Real-time positional variations of the breast/chest wall surface with respect to the reference surface were collected, for all three orthogonal directions. The durations and number of DIBHs needed to complete dose delivery, and DIBH position variations were determined. To evaluate an optimal gating window threshold, smaller tolerances of ±2.5 mm, ±2.0 mm, and ±1.5 mm were simulated. RESULTS: 525 fractions from 33 patients showed that median DIBH duration was 51 s (range: 30-121 s), and median 4 DIBHs per fraction were needed to complete VMAT dose delivery. Median intra-DIBH stability and intrafractional DIBH reproducibility approximated 1.0 mm in each direction. No large differences were found between patients who preferred to perform the DIBH procedure with (n = 21) and without audio-coaching (n = 12). Simulations demonstrated that gating window tolerances could be reduced from ±3.0 mm to ±2.0 mm, without affecting beam-on status. CONCLUSION: Independent of the use of audio-coaching, this study demonstrates that automated surface-guided gating with DIBH and VMAT proved highly efficient. Patients' DIBH performance far exceeded our expectations compared to earlier experiences and literature. Furthermore, gating window tolerances could be reduced.
引用
- Huijskens S, Granton P, Fremeijer K, et al. Clinical practicality and patient performance for surface-guided automated VMAT gating for DIBH breast cancer radiotherapy. Radiother Oncol. 2024. PMID: 38492672. doi:10.1016/j.radonc.2024.110229