三维适形放疗乳腺癌患者中基于表面引导放疗的照射内运动控制阈值确定的计划稳健性分析
本研究对50例乳腺癌患者的适形治疗计划进行六自由度位移稳健性分析,在位移后的CT上重新计算剂量分布并评估靶区与脊髓剂量。通过迭代方法与随机采样,对靶区剂量覆盖施加约束,获得单自由度和六自由度最大允许位移阈值。结果显示靶区剂量覆盖明显下降、脊髓剂量增加,阈值绝对范围为0.8毫米至5毫米和1.4度至5度。
要点
- 靶区剂量覆盖明显下降,脊髓剂量增加,不同自由度和治疗区域稳健性差异大。
- 照射锁骨上淋巴结时,侧向、纵向、翻滚和偏航方向位移的敏感性尤其高。
- 阈值绝对范围为0.8毫米至5毫米和1.4度至5度,可用于基于表面引导放疗的患者运动控制。
核心结论
- 稳健性分析强调评估位置偏差的剂量学影响时需考虑自由度差异和非对称敏感性,并突出旋转位移的重要性,尤其当照射锁骨上淋巴结时。
- 研究引入了确定六自由度位移限值的实用方法,并确定了一套适用于基于表面引导放疗患者运动控制的阈值。
研究设计与样本
- 研究分析了50例乳腺癌患者适形治疗计划对六自由度位移的稳健性,在位移后的CT扫描上重新计算剂量分布,并评估靶区体积和脊髓剂量。
- 通过迭代方法和随机采样,对靶区剂量覆盖施加剂量约束,依次确定单自由度和六自由度位移的最大允许值。
主要发现
- 观察到靶区剂量覆盖明显下降和脊髓剂量增加。
- 治疗计划对不同自由度或治疗区域的稳健性差异很大。
- 照射锁骨上淋巴结时,侧向、纵向、翻滚或偏航方向位移的敏感性特别高,并表现出部分明显的非对称性。
- 阈值表现出类似特性,绝对值范围为0.8毫米至5毫米和1.4度至5度。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The goal of this study was to obtain maximum allowed shift deviations from planning position in six degrees of freedom (DOF), that can serve as threshold values in surface guided radiation therapy (SGRT) of breast cancer patients. METHODS: The robustness of conformal treatment plans of 50 breast cancer patients against 6DOF shifts was investigated. For that, new dose distributions were calculated on shifted computed tomography scans and evaluated with respect to target volume and spinal cord dose. Maximum allowed shift values were identified by imposing dose constraints on the target volume dose coverage for 1DOF, and consecutively, for 6DOF shifts using an iterative approach and random sampling. RESULTS: Substantial decreases in target dose coverage and increases of spinal cord dose were observed. Treatment plans showed highly differing robustness for different DOFs or treated area. The sensitivity was particularly high if clavicular lymph nodes were irradiated, for shifts in lateral, vertical, roll or yaw direction, and showed partly pronounced asymmetries. Threshold values showed similar properties with an absolute value range of 0.8 mm to 5 mm and 1.4° to 5°. CONCLUSION: The robustness analysis emphasized the necessity of taking differences between DOFs and asymmetrical sensitivities into account when evaluating the dosimetric impact of position deviations. It also highlighted the importance of rotational shifts, especially if clavicular lymph nodes were irradiated. A practical approach of determining 6DOF shift limits was introduced and a set of threshold values applicable for SGRT based patient motion control was identified.
引用
- Sauer TO, Stillkrieg W, Ott OJ, et al. Plan robustness analysis for threshold determination of SGRT-based intrafraction motion control in 3DCRT breast cancer radiation therapy. Radiat Oncol. 2023. PMID: 37740237. doi:10.1186/s13014-023-02325-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。