呼吸运动补偿的在线预测:针对MRI引导放射治疗的患者特异性门控控制
本研究基于21例肝癌和10例肺癌患者的2D cine-MR序列,开发并验证一种预测肿瘤运动轨迹跨越目标阈值的二值门控信号在线预测算法。在0.4秒和0.6秒预测窗口下测试三种线性预测器和三种循环神经网络(RNN),并评估仅用前30秒作为预热期的自适应线性回归模型。结果显示自适应线性回归平均门控准确率在肝癌和肺癌患者分别为98.3%和98.0%,门控误差分别为44毫秒和45毫秒,且线性回归在0.6秒和0.4秒预测情况下均优于RNN(P<0.05)。
要点
- 自适应线性回归门控准确率肝癌98.3%、肺癌98.0%,门控误差44毫秒和45毫秒。
- 0.6秒和0.4秒预测下线性回归振幅误差显著小于RNN(P<0.05)。
- 仅用前30秒预热期在线自适应参数即可实现有效门控预测。
核心结论
- 开发了一种功能性在线门控控制方案,采用自适应线性回归,在所有研究指标上比复杂RNN方法更具成本效益且更准确。
研究设计与样本
- 收集21例肝癌和10例肺癌患者的2D cine-MR序列,开发二值门控信号预测算法,预测肿瘤运动轨迹相对目标阈值的跨越时间。
- 测试0.4秒和0.6秒预测窗口,使用三种线性预测器和三种循环神经网络(RNN),系统延迟为0.5秒。
- 评估自适应线性回归模型,仅用前30秒作为预热期,在线预测过程中自适应调整模型参数。
主要发现
- 在0.6秒和0.4秒预测情况下,线性回归均优于其他方法,振幅误差显著小于RNN(P<0.05)。
- 自适应线性回归算法表现最佳,肝癌患者平均门控准确率98.3%、门控误差44毫秒;肺癌患者平均门控准确率98.0%、门控误差45毫秒。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This study aims to validate the effectiveness of linear regression for motion prediction of internal organs or tumors on 2D cine-MR and to present an online gating signal prediction scheme that can improve the accuracy of MR-guided radiotherapy for liver and lung cancer. MATERIALS AND METHODS: We collected 2D cine-MR sequences of 21 liver cancer patients and 10 lung cancer patients to develop a binary gating signal prediction algorithm that forecasts the crossing-time of tumor motion traces relative to the target threshold. Both 0.4 s and 0.6 s prediction windows were tested using three linear predictors and three recurrent neural networks (RNNs), given the system delay of 0.5 s. Furthermore, an adaptive linear regression model was evaluated using only the first 30 s as the burn-in period, during which the model parameters were adapted during the online prediction process. The accuracy of the predicted traces was measured using amplitude metrics (MAE, RMSE, and R2), and in addition, we proposed three temporal metrics, namely crossing error, gating error, and gating accuracy, which are more relevant to the nature of the gating signals. RESULTS: In both 0.6 s and 0.4 s prediction cases, linear regression outperformed other methods, demonstrating significantly smaller amplitude errors compared to the RNNs (P < 0.05). The proposed algorithm with adaptive linear regression had the best performance with an average gating accuracy of 98.3% and 98.0%, a gating error of 44 ms and 45 ms, for liver cancer and lung cancer patients, respectively. CONCLUSION: A functional online gating control scheme was developed with an adaptive linear regression that is both more cost-efficient and accurate than sophisticated RNN based methods in all studied metrics.
引用
- Li Y, Li Z, Zhu J, et al. Online prediction for respiratory movement compensation: a patient-specific gating control for MRI-guided radiotherapy. Radiat Oncol. 2023. PMID: 37697360. doi:10.1186/s13014-023-02341-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。