文献速递乳腺癌放疗放疗物理与技术

乳腺癌每日表面引导放疗中每周锥形束CT的最佳分割与时机

Optimal fractionation and timing of weekly cone-beam CT in daily surface-guided radiotherapy for breast cancer.
PubMed 文献 · Radiat Oncol · 2023-07-05

本研究对23例乳腺切除术后接受常规分割放疗的乳腺癌患者采集每日表面引导放疗(SGRT)与每周锥形束CT(CBCT)图像,通过随机选取1至5次CBCT图像构建16种组合模式,计算不同感兴趣区下的CTV-PTV边界、危及器官剂量及靶区剂量覆盖率。结果显示CTV-PTV边界随CBCT频率增加而缩小,CBCTijkl与CBCT12345接近5mm;作者推荐基于每周CBCT、采用同侧乳腺感兴趣区、每日SGRT联合CBCT频率≥3次的模式,且CBCTijk需在第1、2周进行CBCT。

要点

  • CTV-PTV边界随CBCT频率增加而缩小,CBCTijkl与CBCT12345接近5mm。
  • 同侧乳腺感兴趣区下SGRT误差在前后方向较大,各模式靶区剂量均高于全乳腺感兴趣区(P<0.05)。
  • 作者推荐基于每周CBCT、同侧感兴趣区、每日SGRT联合CBCT频率≥3次的模式,CBCTijk需在第1、2周行CBCT。

核心结论

  • 基于每周CBCT,作者推荐采用同侧感兴趣区、每日SGRT联合CBCT频率≥3次的组合模式,且CBCTijk需在第1、2周进行CBCT。

研究设计与样本

  • 对23例乳腺切除术后接受常规分割放疗的乳腺癌患者采集每日SGRT和每周CBCT图像。
  • 从CBCT图像中随机选取1次(CBCT1)、2次(CBCTij)、3次(CBCTijk)、4次(CBCTijkl)和5次(CBCT12345)图像与SGRT融合,共获得16种模式。
  • 基于不同感兴趣区的SGRT摆位误差计算CTV-PTV边界、危及器官剂量及PTV和CTV的剂量覆盖率(V95%)。
  • 采用Pearson和Spearman方法分析各模式间的剂量相关性,并记录患者特异性参数评估其对剂量的影响。

主要发现

  • CTV-PTV边界随CBCT频率增加而减小,CBCTijkl和CBCT12345的边界接近5mm。
  • 同侧乳腺感兴趣区下SGRT误差在前后方向较大,所有模式的靶区剂量均高于全乳腺感兴趣区(P<0.05)。
  • 同侧感兴趣区下,所有模式间的靶区剂量相关性随CBCT时间间隔增加而增加,随CBCT频率增加先降低后升高,拐点为第5周参与CBCT。
  • CBCT123、CBCT124、CBCT125、CBCTijkl和CBCT12345的剂量偏差最小且无显著差异(P>0.05)。
  • CBCT124与CBCT1234之间、以及(CBCTijkl、CBCT12345)与CBCT125之间在PTV偏差百分比分类判定上一致性极好(Kappa=0.704-0.901)。

关键数值与推荐

  • 作者推荐基于每周CBCT、采用同侧感兴趣区、每日SGRT联合CBCT频率≥3次的模式。
  • CBCTijk需在第1周和第2周进行CBCT。

临床怎么用

  • 作者推荐基于每周CBCT、采用同侧感兴趣区、每日SGRT联合CBCT频率≥3次的模式,且CBCTijk需在第1、2周进行CBCT。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Surface-guided radiotherapy (SGRT) has been demonstrated to be a promising supplement to cone-beam computed tomography (CBCT) in adjuvant breast cancer radiotherapy, but a rational combination mode is lacking in clinical practice. The aim of this study was to explore this mode and investigate its impact on the setup and dose accuracy. METHODS AND MATERIALS: Daily SGRT and weekly CBCT images were acquired for 23 patients with breast cancer who received conventional fractionated radiotherapy after lumpectomy. Sixteen modes were acquired by randomly selecting one (CBCT1), two (CBCTij), three (CBCTijk), four (CBCTijkl), and five (CBCT12345) images from the CBCT images for fusion with the SGRT. The CTV-PTV margins, OAR doses, and dose coverage (V95%) of PTV and CTV was calculated based on SGRT setup errors with different regions of interest (ROIs). Dose correlations between these modalities were investigated using Pearson and Spearman's methods. Patient-specific parameters were recorded to assess their impact on dose. RESULTS: The CTV-PTV margins decreased with increasing CBCT frequencies and were close to 5 mm for CBCTijkl and CBCT12345. For the ipsilateral breast ROI, SGRT errors were larger in the AP direction, and target doses were higher in all modes than in the whole breast ROI (P < 0.05). In the ipsilateral ROI, the target dose correlations between all modes increased with increasing CBCT time intervals, decreased, and then increased with increasing CBCT frequencies, with the inflection point being CBCT participation at week 5. The dose deviations in CBCT123, CBCT124, CBCT125, CBCTijkl, and CBCT12345 were minimal and did not differ significantly (P > 0.05). There was excellent agreement between CBCT124 and CBCT1234, and between (CBCTijkl, CBCT12345) and CBCT125 in determining the classification for the percentage of PTV deviation (Kappa = 0.704-0.901). In addition, there were weak correlations between the patient's Dips_b (ipsilateral breast diameter with bolus) and CTV doses in modes with CBCT participation at week 4 (R = 0.270 to 0.480). CONCLUSIONS: Based on weekly CBCT, these modes with ipsilateral ROI and a combination of daily SGRT and a CBCT frequency of ≥ 3 were recommended, and CBCT was required at weeks 1 and 2 for CBCTijk.