文献速递中枢神经肿瘤放疗放疗物理与技术

胶质母细胞瘤患者在放疗期间佩戴肿瘤治疗电场(TTFields)的分次间摆位不确定度评估

Evaluation of Interfraction Setup Uncertainty of Patients With Glioblastoma Wearing TTFields (Tumor Treating Fields) During Radiation Therapy.
PubMed 文献 · Pract Radiat Oncol · 2023-07-10

本研究收集本中心22例胶质母细胞瘤患者的201幅锥形束CT图像,分为TTFields组与对照组,评估6自由度与3自由度摆位误差及三维矢量幅度。TTFields组与对照组在纵轴方向差异有统计学意义(P<.05),TTFields组纵轴位置误差为-0.51±2.05 mm;作者认为TTFields患者每日图像引导放疗似乎是必要的,对无需图像引导放疗的患者推荐PTV外扩5 mm。

要点

  • TTFields组与对照组在纵轴方向摆位误差差异有统计学意义(P<.05)。
  • TTFields组纵轴(头至足方向)位置误差为-0.51±2.05 mm。
  • 作者认为TTFields患者每日图像引导放疗似乎是必要的,非图像引导放疗者推荐PTV外扩5 mm。

核心结论

  • 放疗期间佩戴TTFields增加了摆位不确定度,尤其在纵轴方向,系统误差为1.40 mm、随机误差为1.28 mm。
  • 作者认为TTFields患者每日图像引导放疗似乎是必要的。
  • 作者推荐对无需图像引导放疗的患者PTV外扩5 mm,以提高治疗的安全性和有效性。

研究设计与样本

  • 本研究收集并分析本中心22例患者的201幅锥形束CT图像。
  • 患者按是否佩戴TTFields分为对照组与TTFields组。
  • 评估6自由度与3自由度摆位误差及三维矢量幅度,并推荐无需图像引导放疗患者的估计PTV外扩边界。

主要发现

  • TTFields组与对照组在纵轴方向差异有统计学意义(P<.05),该结果与TTFields组组内比较一致。
  • TTFields组纵轴(头至足方向)位置误差为-0.51±2.05 mm。
  • TTFields组纵轴系统误差为1.40 mm、随机误差为1.28 mm。

关键数值与推荐

  • 对无需图像引导放疗的患者,推荐PTV外扩5 mm。

临床怎么用

  • 作者认为TTFields患者每日图像引导放疗似乎是必要的。
  • 作者推荐对无需图像引导放疗的TTFields患者PTV外扩5 mm,以提高治疗的安全性和有效性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Tumor treating fields (TTFields) with concurrent radiation therapy (RT) might improve the outcome of patients with newly diagnosed glioblastoma. Several trials, including that conducted in our center, have allowed patients to wear TTFields during RT. We aimed to evaluate the setup uncertainty introduced by TTFields and calculate the planning target volume (PTV) margin for clinical reference. METHODS AND MATERIALS: We collected and analyzed 201 cone beam computed tomography images of 22 patients in our center. Patients with or without TTFields were divided into the control and TTFields groups. We evaluated the setup errors in 6 degrees of freedom and 3 degrees of freedom and the magnitudes in the 3-dimensional vectors. An estimated PTV margin for patients requiring nonimaging-guided RT was recommended. RESULTS: A significant difference was observed in the longitudinal axis between the TTFields and control groups (P < .05). These results were consistent with that of the intragroup comparison of the TTFields group. The position error of the longitudinal axis (from head to feet) was -0.51 ± 2.05 mm in the TTFields group. CONCLUSIONS: Wearing TTFields during RT increased the uncertainty, especially in the longitudinal axis, with a system error of 1.40 mm and a random error of 1.28 mm. Daily image guided RT for TTFields patients seems necessary. However, the recommended expansion margin of the PTV is 5 mm for patients requiring nonimage-guided RT to enhance the safety and efficacy of treatment.

来源:Pract Radiat Oncol,查看原文。版权归原作者所有。

引用

  1. Pan M, Xiao Y, Zhu L, et al. Evaluation of Interfraction Setup Uncertainty of Patients With Glioblastoma Wearing TTFields (Tumor Treating Fields) During Radiation Therapy. Pract Radiat Oncol. 2023. PMID: 37437806. doi:10.1016/j.prro.2023.06.006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。