Halcyon系统中乳腺癌患者基于体表标记激光对准与体表引导放疗的初始摆位方法评估

Evaluation of initial patient setup methods for breast cancer between surface-guided radiation therapy and laser alignment based on skin marking in the Halcyon system.
PubMed 文献 · Radiat Oncol · 2023-04-04

本研究回顾性分析228个治疗分次,比较Halcyon系统中体表引导放疗(SGRT)与基于体表标记的激光对准两种初始摆位方法的精度与效率。SGRT组在垂直和横向平移误差及俯仰、翻滚旋转误差上数值更低,平均总摆位时间SGRT组为314秒,无SGRT组为331秒。结论认为SGRT提高了Halcyon系统乳腺癌初始摆位的精度和效率,而该系统在纠正旋转偏移方面存在局限。

要点

  • SGRT组平移误差垂直1.9±1.2 mm、纵向2.9±2.1 mm、横向1.9±0.7 mm,激光对准组分别为2.7±1.6 mm、2.0±1.2 mm、2.1±1.0 mm。
  • SGRT组旋转误差俯仰0.30±0.22°、偏航0.51±0.26°、翻滚0.19±0.13°,激光对准组分别为0.32±0.30°、0.51±0.24°、0.29±0.22°。
  • 平均总摆位时间SGRT组314秒,无SGRT组331秒。

核心结论

  • 使用SGRT提高了Halcyon系统中乳腺癌患者初始摆位的精度和效率,而该系统在纠正旋转偏移方面存在局限。

研究设计与样本

  • 回顾性研究,共分析228个治疗分次,比较两种初始摆位方法:基于体表标记的激光对准和体表引导放疗(SGRT)。
  • 使用自制乳腺模体评估SGRT系统的残余旋转误差精度。
  • 通过记录与验证系统获取的参考CT与每日千伏级锥形束CT之间在垂直、纵向和横向的床位置差异分析残余平移误差。
  • 基于Velocity软件对两幅图像进行自动刚性配准计算残余旋转误差(俯仰、偏航、翻滚)。
  • 分析总摆位时间(初始摆位时间与成像时间之和)以评估SGRT每日摆位的效率。

主要发现

  • 使用自制乳腺模体测得的平均残余旋转误差俯仰为0.14°、翻滚为0.13°、偏航为0.29°。
  • 基于体表标记的激光对准组平均床位置差异:垂直2.7±1.6 mm、纵向2.0±1.2 mm、横向2.1±1.0 mm。
  • SGRT组平均床位置差异:垂直1.9±1.2 mm、纵向2.9±2.1 mm、横向1.9±0.7 mm。
  • 无SGRT时旋转误差俯仰0.32±0.30°、偏航0.51±0.24°、翻滚0.29±0.22°;SGRT时分别为0.30±0.22°、0.51±0.26°、0.19±0.13°。
  • 考虑初始摆位时间和成像时间的平均总摆位时间:有SGRT为314秒,无SGRT为331秒。

关键数值与推荐

  • SGRT组垂直平移误差1.9±1.2 mm低于激光对准组2.7±1.6 mm,横向1.9±0.7 mm低于2.1±1.0 mm,纵向2.9±2.1 mm高于2.0±1.2 mm。
  • SGRT组翻滚旋转误差0.19±0.13°低于激光对准组0.29±0.22°,俯仰0.30±0.22°低于0.32±0.30°,偏航0.51±0.26°与0.51±0.24°相近。
  • 平均总摆位时间SGRT组314秒短于无SGRT组331秒。

临床怎么用

  • 原文结论未据此给出具体临床推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: This study was conducted to evaluate the efficiency and accuracy of the daily patient setup for breast cancer patients by applying surface-guided radiation therapy (SGRT) using the Halcyon system instead of conventional laser alignment based on the skin marking method. METHODS AND MATERIALS: We retrospectively investigated 228 treatment fractions using two different initial patient setup methods. The accuracy of the residual rotational error of the SGRT system was evaluated by using an in-house breast phantom. The residual translational error was analyzed using the couch position difference in the vertical, longitudinal, and lateral directions between the reference computed tomography and daily kilo-voltage cone beam computed tomography acquired from the record and verification system. The residual rotational error (pitch, yaw, and roll) was also calculated using an auto rigid registration between the two images based on Velocity. The total setup time, which combined the initial setup time and imaging time, was analyzed to evaluate the efficiency of the daily patient setup for SGRT. RESULTS: The average residual rotational errors using the in-house fabricated breast phantom for pitch, roll, and yaw were 0.14°, 0.13°, and 0.29°, respectively. The average differences in the couch positions for laser alignment based on the skin marking method were 2.7 ± 1.6 mm, 2.0 ± 1.2 mm, and 2.1 ± 1.0 mm for the vertical, longitudinal, and lateral directions, respectively. For SGRT, the average differences in the couch positions were 1.9 ± 1.2 mm, 2.9 ± 2.1 mm, and 1.9 ± 0.7 mm for the vertical, longitudinal, and lateral directions, respectively. The rotational errors for pitch, yaw, and roll without the surface-guided radiation therapy approach were 0.32 ± 0.30°, 0.51 ± 0.24°, and 0.29 ± 0.22°, respectively. For SGRT, the rotational errors were 0.30 ± 0.22°, 0.51 ± 0.26°, and 0.19 ± 0.13°, respectively. The average total setup times considering both the initial setup time and imaging time were 314 s and 331 s, respectively, with and without SGRT. CONCLUSION: We demonstrated that using SGRT improves the accuracy and efficiency of initial patient setups in breast cancer patients using the Halcyon system, which has limitations in correcting the rotational offset.