四维锥形束CT在肺癌放疗中的应用
本研究对67例肺癌患者放疗期间应用四维锥形束CT(4D CBCT)评估肿瘤体积反应、运动和中心坐标,并比较4D CBCT与3D CBCT在不同配准方法下的差异;41%患者观察到肿瘤体积反应,平均体积缩小41.7%,中位出现时间19天;仅基于骨配准时41.8%的病例出现部分脱靶。
要点
- 41%(28/67)患者放疗期间出现肿瘤体积反应,平均缩小41.7%,中位时间19天。
- 16例肿瘤运动明显,平均0.52 cm(范围0.22–1.34 cm)。
- 仅基于骨配准时41.8%病例部分脱靶;运动≤0.5 cm脱靶率19.0%,>0.5 cm为52.2%。
核心结论
- 肺癌患者肺内病灶的肿瘤体积和运动在放疗第三周变化显著。
- 4D CBCT对于孤立病灶、无相对解剖结构参考或靠近膈肌的病灶可能更有优势。
- 基于平均密度投影的灰度配准是可行的。
研究设计与样本
- 对67例符合条件的肺癌患者放疗期间应用4D CBCT评估肿瘤体积反应、运动和中心坐标。
- 比较4D CBCT与3D CBCT在不同配准方法下的差异。
主要发现
- 41%(28/67)患者治疗期间观察到肿瘤体积反应,平均体积缩小41.7%,中位出现时间19天。
- 16例患者肿瘤运动明显,平均0.52 cm(范围0.22–1.34 cm);靠近膈肌的6个肿瘤中3个运动范围为0.28–0.66 cm。
- 基于平均密度投影的灰度值配准仍可获得与4D灰度值配准接近的结果。
- 仅基于骨配准时,41.8%的病例治疗中出现部分脱靶。
- 肿瘤运动≤0.5 cm时脱靶率为19.0%,运动>0.5 cm时脱靶率为52.2%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: This study explored the application of four-dimensional cone beam computed tomography (4D CBCT) in lung cancer patients, seeking to improve the accuracy of radiotherapy and to establish a uniform protocol for the application of 4D CBCT in radiotherapy for lung cancer. METHODS: 4D CBCT was applied to evaluate tumor volume response (TVR), motion, and center coordinates during radiotherapy in 67 eligible individuals with lung cancer diagnoses. The differences between 4D CBCT and 3D CBCT in different registration methods were compared. RESULTS: TVR was observed during treatment in 41% of patients (28/67), with a mean volume reduction of 41.7% and a median time to TVR of 19 days. Tumor motion was obvious in 16 patients, with a mean value of 0.52 cm (0.22 to 1.34 cm), and in 3 of 6 tumors close to the diaphragm (0.28 to 0.66 cm). Gray value registration based on mean density projection could still achieve close results to the 4D gray value registration. However, when the registration was based on bone alone, partial off-targeting occurred in the treatment in 41.8% of cases. The off-target rate was 19.0% when the tumor motion was ≤ 0.5 cm and 52.2% when the motion was > 0.5 cm. CONCLUSION: Tumor volume and motion of intrapulmonary lesions in individuals diagnosed with lung cancer varied significantly in the third week of radiotherapy. 4D CBCT may be more advantageous for isolated lesions without reference to relative anatomical structures or those near the diaphragm. Grayscale registration based on mean density projection is feasible.
引用
- Abulimiti M, Yang X, Li M, et al. Application of four-dimensional cone beam computed tomography in lung cancer radiotherapy. Radiat Oncol. 2023. PMID: 37069641. doi:10.1186/s13014-023-02259-8
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。