加速部分乳腺照射中表面成像与皮肤纹身患者摆位准确性和效率的比较

Accuracy and Efficiency of Patient Setup Using Surface Imaging versus Skin Tattoos for Accelerated Partial Breast Irradiation.
PubMed 文献 · Adv Radiat Oncol · 2023-01-26

本研究在43例接受加速部分乳腺照射的患者中交替使用传统纹身摆位(TTB)与基于AlignRT的表面成像无纹身摆位(ART),共分析356个治疗分次,以kV影像配准手术夹为基准。ART在纵向平移和总位移幅度上优于TTB,各平移轴方差更窄,总室内时间和摆位时间更短,旋转摆位与TTB无显著差异。作者认为AlignRT无纹身摆位可能足够准确和快速,或可替代表面纹身。

要点

  • ART组总位移幅度中位数0.59 cm,低于TTB组的0.80 cm(P < .001)。
  • ART组各平移轴方差显著窄于TTB组(垂直P ≤ .001,横向P = .001,纵向P = .005)。
  • ART组总室内时间中位数15.42分钟,短于TTB组的17.25分钟(P = .008)。

核心结论

  • 作者认为,使用AlignRT的无纹身摆位方法可能足够准确和快速,或可替代加速部分乳腺照射患者的表面纹身。
  • 作者指出,需要更大队列的进一步分析来确定基于纹身的方法能否被无创表面成像替代。

研究设计与样本

  • 患者接受加速部分乳腺照射,传统纹身摆位(TTB)与基于AlignRT的表面成像无纹身摆位(ART)每日交替进行。
  • 初始摆位后通过每日kV成像验证位置,以手术夹配准作为基准真值。
  • 共分析43例接受加速部分乳腺照射的患者和356个治疗分次(TTB组174个分次,ART组182个分次)。
  • 统计分析使用Wilcoxon符号秩检验和Pitman-Morgan方差检验。

主要发现

  • ART组平移中位绝对位移:垂直0.31 cm(范围0.08-0.82)、横向0.23 cm(0.05-0.86)、纵向0.26 cm(0.02-0.72);TTB组对应值分别为0.34 cm(0.05-1.98)、0.31 cm(0.09-1.84)、0.34 cm(0.08-1.25)。
  • ART组总位移幅度中位数0.59(0.30-1.31),TTB组为0.80(0.27-2.13);ART在总位移幅度上优于TTB(P < .001)。
  • ART与TTB在平移位移上除纵向方向外无统计学差异(P分别为.154、.059和.021),各平移轴方差ART均显著窄于TTB(垂直P ≤ .001,横向P = .001,纵向P = .005)。
  • ART组旋转中位绝对位移:旋转0.64°(范围0.00-1.90)、滚动0.65°(0.05-2.90)、俯仰0.30°(0.00-1.50);TTB组对应值分别为0.80°(0.00-2.50)、0.64°(0.00-3.00)、0.46°(0.00-2.90);旋转位移两组无统计学差异(P分别为.868、.236和.079),俯仰方差ART低于TTB(P = .009)。
  • ART组总室内时间中位数15.42分钟 vs TTB组17.25分钟(P = .008),摆位时间中位数11.12分钟 vs 13.00分钟(P = .001);ART摆位时间分布更窄,长时离群值更少。

临床怎么用

  • 作者认为,使用AlignRT的无纹身摆位方法可能足够准确和快速,或可替代加速部分乳腺照射患者的表面纹身。
  • 作者指出,需要更大队列的进一步分析来确定基于纹身的方法能否被无创表面成像替代。

局限性

  • 作者指出,需要更大队列的进一步分析来确定基于纹身的方法能否被无创表面成像替代。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Skin tattoos represent the standard approach for surface alignment and setup of breast cancer radiation therapy, yet permanent skin markings contribute to adverse cosmesis and patient dissatisfaction. With the advent of contemporary surface-imaging technology, we evaluated setup accuracy and timing between "tattoo-less" and traditional tattoo-based setup techniques. METHODS AND MATERIALS: Patients receiving accelerated partial breast irradiation (APBI) underwent traditional tattoo-based setup (TTB), alternating daily with a tattoo-less setup via surface imaging using AlignRT (ART). Following initial setup, position was verified via daily kV imaging, with matching on surgical clips representing ground truth. Translational shifts (TS) and rotational shifts (RS) were ascertained, as were setup time and total in-room time. Statistical analyses used the Wilcoxon signed rank test and Pitman-Morgan variance test. RESULTS: A total of 43 patients receiving APBI and 356 treatment fractions were analyzed (174 TTB fractions and 182 using ART). For tattoo-less setup via ART, the median absolute TS were 0.31 cm in the vertical (range, 0.08-0.82), 0.23 cm in the lateral (0.05-0.86), and 0.26 cm in the longitudinal (0.02-0.72) axes. For TTB setup, the corresponding median TS were 0.34 cm (0.05-1.98), 0.31 cm (0.09-1.84), and 0.34 cm (0.08-1.25), respectively. The median magnitude shifts were 0.59 (0.30-1.31) for ART and 0.80 (0.27-2.13) for TTB. ART was not statistically distinguishable from TTB in terms of TS, except in the longitudinal direction (P = .154, .059, and .021, respectively), and was superior to TTB for magnitude shift (P < .001). The variance of each TS variable was significantly narrower for ART compared with TTB (P ≤ .001 vertical, P = .001 lateral, P = .005 longitudinal). The median absolute RS for ART was 0.64° rotation (range, 0.00-1.90), 0.65° roll (0.05-2.90), and 0.30° pitch (0.00-1.50). The corresponding median RS for TTB were 0.80° (0.00-2.50), 0.64° (0.00-3.00), and 0.46° (0.00-2.90), respectively. ART setup was not statistically different from TTB in terms of RS (P = .868, .236, and .079, respectively). ART showed lower variance than TTB in terms of pitch (P = .009). The median total in-room time was shorter for ART than TTB (15.42 vs 17.25 minutes; P = .008), as was the median setup time (11.12 vs 13.00 minutes; P = .001). Moreover, ART had a narrower distribution of setup time with fewer lengthy outliers versus TTB. CONCLUSIONS: These findings suggest that a tattoo-less setup approach with AlignRT may be sufficiently accurate and expeditious to supplant surface tattoos for patients receiving APBI. Further analyses with larger cohorts will determine whether tattoo-based approaches can be replaced by noninvasive surface imaging.