文献速递放疗物理与技术

利用机器学习表征治疗参数与复杂性之间的相互作用及其对IROC IMRT体模性能的影响

Characterizing the interplay of treatment parameters and complexity and their impact on performance on an IROC IMRT phantom using machine learning.
PubMed 文献 · Radiother Oncol · 2023-02-24

该研究回顾性分析了2012至2020年间1542次IROC头颈部IMRT体模照射,基于输出参数、复杂性指标和治疗参数,采用单因素分析和随机森林建模。体模平均通过率为92%且未随时间显著改善,步进式照射技术通过率显著更低,计划复杂性随时间显著增加,基于孔径的复杂性指标(除MCS外)均与失败概率相关。

要点

  • 体模平均通过率为92%,未随时间显著改善。
  • 步进式照射技术通过率显著更低,并显著影响其他治疗参数的通过率。
  • 基于孔径的复杂性指标(除MCS外)均与失败概率相关,但其重要性取决于具体治疗参数。

核心结论

  • 复杂性指标对照射失败有强预测作用,但其预测作用依赖于具体的治疗参数。
  • 使用单一复杂性指标不足以监测治疗计划的所有方面。

研究设计与样本

  • 回顾性分析2012至2020年间1542次IROC头颈部IMRT体模照射。
  • 体模包含两个靶区和一个危及器官,使用TLD测量点剂量、胶片测量二维剂量。
  • 基于输出参数、复杂性指标和治疗参数进行单因素分析,并使用随机森林建模预测输出参数并确定变量重要性。

主要发现

  • 体模平均通过率为92%,且未随时间显著改善。
  • 步进式照射技术的通过率显著更低,并显著影响其他治疗参数的通过率。
  • 计划复杂性随时间显著增加,所有基于孔径的复杂性指标(除MCS外)均与失败概率相关。
  • 随机森林预测失败在未用于模型训练的保留测试数据上准确率为98%。
  • 复杂性指标是最重要的贡献因素,但具体哪个指标取决于照射时使用的治疗参数集。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM OF THE STUDY: To elucidate the important factors and their interplay that drive performance on IMRT phantoms from the Imaging and Radiation Oncology Core (IROC). METHODS: IROC's IMRT head and neck phantom contains two targets and an organ at risk. Point and 2D dose are measured by TLDs and film, respectively. 1,542 irradiations between 2012-2020 were retrospectively analyzed based on output parameters, complexity metrics, and treatment parameters. Univariate analysis compared parameters based on pass/fail, and random forest modeling was used to predict output parameters and determine the underlying importance of the variables. RESULTS: The average phantom pass rate was 92% and has not significantly improved over time. The step-and-shoot irradiation technique had significantly lower pass rates that significantly affected other treatment parameters' pass rates. The complexity of plans has significantly increased with time, and all aperture-based complexity metrics (except MCS) were associated with the probability of failure. Random forest-based prediction of failure had an accuracy of 98% on held-out test data not used in model training. While complexity metrics were the most important contributors, the specific metric depended on the set of treatment parameters used during the irradiation. CONCLUSION: With the prevalence of errors in radiotherapy, understanding which parameters affect treatment delivery is vital to improve patient treatment. Complexity metrics were strongly predictive of irradiation failure; however, they are dependent on the specific treatment parameters. In addition, the use of one complexity metric is insufficient to monitor all aspects of the treatment plan.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Mehrens H, Molineu A, Hernandez N, et al. Characterizing the interplay of treatment parameters and complexity and their impact on performance on an IROC IMRT phantom using machine learning. Radiother Oncol. 2023. PMID: 36841341. doi:10.1016/j.radonc.2023.109577

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。