ESTRO-ACROP指南:基于共识专家意见的在线MRI引导放疗系统质量保证与医学物理调试
该共识专家意见由十名医学物理师和两名放射肿瘤学家参与,通过三轮投票定义了在线MRI引导放疗(oMRgRT)系统相较于在线CT引导放疗系统需额外考虑的24项质量保证测试及11项建设、工作流程和培训任务,并就多数测试的重要性、推荐频率和适用性达成共识,同时确定8个与工作流程定义相关的调试重要方面。
要点
- 共识别24项oMRgRT系统质量保证测试,与在线CT引导放疗系统不同。
- 另收集11项与建设、工作流程开发和培训相关的任务与方面。
- 多数测试的重要性、推荐频率及对两种获批系统的适用性达成共识。
核心结论
- 制定了一套oMRgRT系统质量保证与调试方案,以支持医学物理师为安全操作此类系统做好准备。
研究设计与样本
- 共识专家意见:十名医学物理师和两名放射肿瘤学家参与,经三轮投票就质量保证测试的重要性、推荐频率、对两种获批临床使用的oMRgRT系统的适用性以及医学物理调试需考虑方面的重要性达成共识。
主要发现
- 识别出24项在oMRgRT系统调试和常规质量保证中可能重要的测试,这些测试区别于在线CT引导放疗系统。
- 额外收集了11项与建设、工作流程开发和培训相关的任务与方面。
- 多数测试在其重要性、推荐频率和对两种获批系统的适用性上达成共识。
- 另发现8个主要与工作流程定义相关的方面在调试期间也属重要。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The goal of this consensus expert opinion was to define quality assurance (QA) tests for online magnetic resonance image (MRI) guided radiotherapy (oMRgRT) systems and to define the important medical physics aspects for installation and commissioning of an oMRgRT system. MATERIALS AND METHODS: Ten medical physicists and two radiation oncologists experienced in oMRgRT participated in the survey. In the first round of the consensus expert opinion, ideas on QA and commissioning were collected. Only tests and aspects different from commissioning of a CT guided radiotherapy (RT) system were considered. In the following two rounds all twelve participants voted on the importance of the QA tests, their recommended frequency and their suitability for the two oMRgRT systems approved for clinical use as well as on the importance of the aspects to consider during medical physics commissioning. RESULTS: Twenty-four QA tests were identified which are potentially important during commissioning and routine QA on oMRgRT systems compared to online CT guided RT systems. An additional eleven tasks and aspects related to construction, workflow development and training were collected. Consensus was found for most tests on their importance, their recommended frequency and their suitability for the two approved systems. In addition, eight aspects mostly related to the definition of workflows were also found to be important during commissioning. CONCLUSIONS: A program for QA and commissioning of oMRgRT systems was developed to support medical physicists to prepare for safe handling of such systems.
引用
- Tanadini-Lang S, Budgell G, Bohoudi O, et al. An ESTRO-ACROP guideline on quality assurance and medical physics commissioning of online MRI guided radiotherapy systems based on a consensus expert opinion. Radiother Oncol. 2023. PMID: 36736592. doi:10.1016/j.radonc.2023.109504
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。