面向高精度MRI引导颅内放疗的经验性计划靶区建模

Empirical planning target volume modeling for high precision MRI guided intracranial radiotherapy.
PubMed 文献 · Clin Transl Radiat Oncol · 2023-01-16

本研究纳入66例(49例原发脑肿瘤、17例术后切除腔)在1.5T MR-linac上治疗的1329个分次,量化仅平移摆位与分次内运动的不确定性,建立基于人群的PTV模型。覆盖95%患者、95%分次中98%的CTVfx需3 mm PTV外扩;外扩从4 mm减至3 mm可使视交叉与PTV重叠比例最多降低23.5%。作者据此临床采用3 mm PTV外扩用于常规分割颅内患者。

要点

  • 覆盖95%患者、95%分次中98%的CTVfx需3 mm PTV外扩。
  • 外扩从4 mm减至3 mm使视交叉与PTV重叠比例最多降低23.5%。
  • 作者据此临床采用3 mm PTV外扩用于常规分割颅内患者。

核心结论

  • 作者建立了基于证据的PTV模型,并据此临床采用3 mm PTV外扩用于常规分割颅内患者。

研究设计与样本

  • 纳入66例患者(49例原发脑肿瘤、17例术后切除腔),在1.5T MR-linac上共治疗1329个分次。
  • 每个分次仅平移融合在线T1 MRI与计划图像完成摆位,离线独立重复融合并计入旋转,将六自由度差异应用于计划CTV得到CTVfx。
  • 建立按CTVfx体积覆盖、分次比例和患者比例参数化的PTV模型;在412个分次子集中量化分次内运动。

主要发现

  • 左右/前后/头脚方向旋转融合差异均值±标准差分别为0.1±0.8/0.1±0.8/-0.2±0.9°。
  • 覆盖95%患者、95%分次中98%的CTVfx需3 mm PTV外扩。
  • 外扩从4 mm减至3 mm使视交叉与PTV重叠比例最多降低23.5%。

关键数值与推荐

  • 覆盖95%患者、95%分次中98%的CTVfx的PTV外扩为3 mm。
  • 外扩从4 mm减至3 mm使视交叉与PTV重叠比例最多降低23.5%。

临床怎么用

  • 作者表示已根据该模型对常规分割颅内患者临床采用3 mm PTV外扩。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Magnetic resonance image-guided radiotherapy for intracranial indications is a promising advance; however, uncertainties remain for both target localization after translation-only MR setup and intrafraction motion. This investigation quantified these uncertainties and developed a population-based planning target volume (PTV) model to explore target and organ-at-risk (OAR) volumetric coverage tradeoffs. METHODS: Sixty-six patients, 49 with a primary brain tumor and 17 with a post-surgical resection cavity, treated on a 1.5T-based MR-linac across 1329 fractions were included. At each fraction, patients were setup by translation-only fusion of the online T1 MRI to the planning image. Each fusion was independently repeated offline accounting for rotations. The six degree-of-freedom difference between fusions was applied to transform the planning CTV at each fraction (CTVfx). A PTV model parameterized by volumetric CTVfx coverage, proportion of fractions, and proportion of patients was developed. Intrafraction motion was quantified in a 412 fraction subset as the fusion difference between post- and pre-irradiation T1 MRIs. RESULTS: For the left-right/anterior-posterior/superior-inferior axes, mean ± SD of the rotational fusion differences were 0.1 ± 0.8/0.1 ± 0.8/-0.2 ± 0.9°. Covering 98 % of the CTVfx in 95 % of fractions in 95 % of patients required a 3 mm PTV margin. Margin reduction decreased PTV-OAR overlap; for example, the proportion of optic chiasm overlapped by the PTV was reduced up to 23.5 % by margin reduction from 4 mm to 3 mm. CONCLUSIONS: An evidence-based PTV model was developed for brain cancer patients treated on the MR-linac. Informed by this model, we have clinically adopted a 3 mm PTV margin for conventionally fractionated intracranial patients.