利用TCP/NTCP模型评估前列腺癌放疗中CBCT软组织匹配可能存在的观察者间变异的影响
该回顾性研究分析12例接受 elective 淋巴结照射的前列腺癌患者的156幅CBCT图像,模拟观察者间变异(相对于前列腺对齐结果假设2 mm床移位,共27种可能位置),用可变形图像配准合成的CT重新计算剂量,并以TCP和NTCP评估影响。结果显示TCP无显著差异,直肠NTCP受前后方向床移位影响较大,膀胱NTCP相对可忽略。
要点
- 前列腺对齐与模拟位置之间TCP无显著差异(0.944±0.003 vs 0.945±0.003,P=0.117)。
- 直肠平均NTCP在各模拟位置间为5.16%至7.29%,受前后方向床移位高度影响。
- 膀胱平均NTCP在各模拟位置间为0.75%至1.12%,影响相对可忽略。
核心结论
- CBCT软组织匹配的观察者间变异对直肠NTCP影响较大,而对靶区TCP影响较小。
研究设计与样本
- 回顾性分析12例接受 elective 淋巴结照射的前列腺癌患者的156幅CBCT图像。
主要发现
- 前列腺对齐与模拟位置之间TCP无显著差异(0.944±0.003 vs 0.945±0.003,P=0.117)。
- 直肠平均NTCP在各模拟位置间为5.16%至7.29%,受前后方向床移位高度影响。
- 膀胱平均NTCP在各模拟位置间为0.75%至1.12%,影响相对可忽略。
临床怎么用
- 原文结论指出本研究为每日CBCT图像引导及前列腺-直肠界面匹配流程提供了理论解释。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Prostate alignment is subject to interobserver variability in cone-beam CT (CBCT)-based soft-tissue matching. This study aims to analyze the impact of possible interobserver variability in CBCT-based soft-tissue matching for prostate cancer radiotherapy. METHODS: Retrospective data, consisting of 156 CBCT images from twelve prostate cancer patients with elective nodal irradiation were analyzed in this study. To simulate possible interobserver variability, couch shifts of 2 mm relative to the resulting patient position of prostate alignment were assumed as potential patient positions (27 possibilities). For each CBCT, the doses of the potential patient positions were re-calculated using deformable image registration-based synthetic CT. The impact of the simulated interobserver variability was evaluated using tumor control probabilities (TCPs) and normal tissue complication probabilities (NTCPs). RESULTS: No significant differences in TCPs were found between prostate alignment and potential patient positions (0.944 ± 0.003 vs 0.945 ± 0.003, P = 0.117). The average NTCPs of the rectum ranged from 5.16 to 7.29 (%) among the potential patient positions and were highly influenced by the couch shift in the anterior-posterior direction. In contrast, the average NTCPs of the bladder ranged from 0.75 to 1.12 (%) among the potential patient positions and were relatively negligible. CONCLUSIONS: The NTCPs of the rectum, rather than the TCPs of the target, were highly influenced by the interobserver variability in CBCT-based soft-tissue matching. This study provides a theoretical explanation for daily CBCT-based image guidance and the prostate-rectum interface matching procedure. TRIAL REGISTRATION: Not applicable.
引用
- Zhang X, Wang X, Li X, et al. Evaluating the impact of possible interobserver variability in CBCT-based soft-tissue matching using TCP/NTCP models for prostate cancer radiotherapy. Radiat Oncol. 2022. PMID: 35365155. doi:10.1186/s13014-022-02034-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。