患者特异性植入物与CAD/CAM夹板在正颌手术中准确性的随机临床试验
这项随机对照试验纳入64例需正颌手术患者,比较患者特异性植入物与CAD/CAM手术夹板在上颌骨复位中的准确性。分析27例植入物组和31例夹板组,上颌骨位置偏差分别为1.41±0.58 mm和2.20±0.94 mm,组间差异显著(p<0.001),表明患者特异性植入物复位更准确。
要点
- 患者特异性植入物组上颌骨位置偏差1.41±0.58 mm,优于夹板组2.20±0.94 mm(p<0.001)。
- 植入物组最大平移偏差为前后向1.02±0.66 mm,最大方向偏差为俯仰1.85±1.42度。
- 夹板组最大平移偏差为内外侧1.23±0.93 mm,最大方向偏差为俯仰1.72±1.56度。
核心结论
- 正颌手术中使用患者特异性植入物比CAD/CAM手术夹板能获得更准确的上颌骨位置。
研究设计与样本
- 随机对照试验,纳入64例需正颌手术患者,随机分配至患者特异性植入物组或CAD/CAM夹板组。
- 最终分析患者特异性植入物组27例、夹板组31例,通过比较虚拟计划与实际结果评估准确性。
主要发现
- 上颌骨质心位置偏差在患者特异性植入物组为1.41±0.58 mm,夹板组为2.20±0.94 mm,组间差异显著(p<0.001)。
- 患者特异性植入物组最大平移偏差为前后向1.02±0.66 mm,最大方向偏差为俯仰1.85±1.42度。
- 夹板组最大平移偏差为内外侧1.23±0.93 mm,最大方向偏差为俯仰1.72±1.56度。
临床怎么用
- 对于需正颌手术行上颌骨复位的患者,可优先考虑使用患者特异性植入物以提高位置准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The maxilla position is essential for the aesthetic and functional outcomes of orthognathic surgery. Previous studies demonstrated the advantages of patient-specific implants in orthognathic surgery. However, more data are needed to confirm the superiority of patient-specific implants over surgical splints created with computer-aided design/computer-aided manufacturing (CAD/CAM). This randomized controlled trial aimed to compare the accuracy of patient-specific implants and CAD/CAM splints for maxilla repositioning in orthognathic surgery. METHODS: Patients (n = 64) who required orthognathic surgery were randomly assigned to use either patient-specific implants (patient-specific implant group) or CAD/CAM surgical splints (splint group) to reposition the maxilla. The outcome evaluation was completed by comparing virtual plans with actual results. The primary outcome was the discrepancies of the centroid position of the maxilla. Other translation and orientation discrepancies of the maxilla were also assessed. RESULTS: The authors analyzed 27 patients in the patient-specific implant group and 31 in the splint group. The maxilla position discrepancy was 1.41 ± 0.58 mm in the patient-specific implant group and 2.20 ± 0.94 mm in the splint group; the between-group difference was significant (p < 0.001). For the patient-specific implant group, the largest translation discrepancy was 1.02 ± 0.66 mm in the anteroposterior direction, and the largest orientation discrepancy was 1.85 ± 1.42 degrees in pitch. For the splint group, the largest translation discrepancy was 1.23 ± 0.93 mm in the mediolateral direction, and the largest orientation discrepancy was 1.72 ± 1.56 degrees in pitch. CONCLUSION: The result showed that using patient-specific implants in orthognathic surgery resulted in a more accurate maxilla position than CAD/CAM surgical splints. CLINICAL QUESTION/LEVEL OF EVIDENCE: Therapeutic, I.
引用
- Li B, Wei H, Jiang T, et al. Randomized Clinical Trial of the Accuracy of Patient-Specific Implants versus CAD/CAM Splints in Orthognathic Surgery. Plast Reconstr Surg. 2021. PMID: 34705785. doi:10.1097/PRS.0000000000008427
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。