Intra.Ox近红外光谱仪测量皮瓣氧合变化与坏死相关性的临床前大鼠模型研究

The Intra.Ox Near-Infrared Spectrometer Measures Variations in Flap Oxygenation That Correlate to Flap Necrosis in a Preclinical Rodent Model.
PubMed 文献 · Plast Reconstr Surg · 2021-01-01

本研究在大鼠背部随意皮瓣模型中,使用Intra.Ox近红外光谱仪术中测量10个标准化位点的组织氧饱和度,并与术后第7天全层坏死进行关联分析。结果显示氧饱和度与距蒂部距离呈强负相关(r=-0.798),坏死组织平均氧饱和度32%显著低于非坏死组织59%(p<0.001),ROC曲线下面积为0.969。该技术可识别高危组织,为预防皮瓣坏死的研究提供方向。

要点

  • Intra.Ox可检测皮瓣组织氧饱和度差异,与坏死风险相关。
  • 坏死组织氧饱和度平均32%,非坏死组织59%(p<0.001)。
  • 氧饱和度与距蒂部距离强负相关(r=-0.798),ROC曲线下面积0.969。

核心结论

  • Intra.Ox检测到与皮瓣坏死风险相关的组织氧饱和度显著差异,可用于识别高危组织。

研究设计与样本

  • 在8只雄性Sprague-Dawley大鼠背部制备10×3 cm随意皮瓣,术中测量10个标准化位点氧饱和度。
  • 术后第7天评估全层坏死,采用卡方检验、单因素方差分析和ROC曲线分析。

主要发现

  • 组织氧饱和度与距皮瓣蒂部距离呈强负相关(r=-0.798)。
  • 坏死组织平均氧饱和度为32%,非坏死组织为59%(p<0.001)。
  • ROC曲线下面积为0.969。
  • 预测坏死的氧饱和度截断值在46%和54%时分别达到100%特异度和100%敏感度。

关键数值与推荐

  • 氧饱和度截断值46%时预测坏死特异度100%,54%时敏感度100%。

临床怎么用

  • Intra.Ox可术中连续评估皮瓣灌注,帮助客观识别高危组织并指导治疗决策。

局限性

  • 本研究为临床前大鼠模型初步研究,样本量小,结论需进一步临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Mastectomy flap necrosis affects 7 to 40 percent of patients undergoing immediate breast reconstruction, with many cases resulting in infection and/or explantation. The Intra.Ox near-infrared spectrometer is a novel device that assesses tissue perfusion by measuring the interactions of light with oxygenated and deoxygenated hemoglobin. This handheld device facilitates serial flap perfusion assessment and may objectively identify at-risk tissues and guide evidence-based treatment algorithms. In this preliminary study, we hypothesized that the Intra.Ox spectrometer detects differences in tissue oxygenation that correlate to tissue necrosis. METHODS: Dorsal, random-pattern flaps measuring 10 × 3 cm were raised in eight male Sprague-Dawley rats. Intraoperative tissue oxygen saturation was measured using Intra.Ox in 10 standardized locations. On postoperative day 7, the skin flaps were evaluated for full-thickness necrosis. Data were analyzed using the chi-square test and one-way analysis of variance. A receiver operating characteristic curve assessed the accuracy of intraoperative tissue oxygenation in predicting the risk of flap necrosis. RESULTS: Tissue oxygen saturation exhibited a strong negative correlation to distance from the flap pedicle (r = -0.798). Oxygen saturation in tissue that developed necrosis averaged 32 percent, compared to 59 percent in tissues that did not (p < 0.001). The area under the receiver operating characteristic curve was 0.969. Post hoc oxygen saturation cutoffs with 100 percent specificity and sensitivity in predicting necrosis were identified at 46 percent and 54 percent, respectively. CONCLUSIONS: Intra.Ox detects significant differences in tissue oxygenation saturation that are associated with the risk for flap necrosis. This technology can be used to identify at-risk tissues and represents an avenue for research aimed at preventing flap necrosis.