文献速递乳房整形与再造

AeroForm组织扩张器早期意外充气一例

AeroForm Tissue Expander: A Case of Early Accidental Inflation.
PubMed 文献 · Ann Plast Surg · 2021-01-01

本文报道一例57岁导管原位癌女性在双侧保留乳头乳房切除术后置入V2.5 AeroForm二氧化碳组织扩张器,进入磁共振检查室时双侧扩张器储气阀突然激活,释放气体至800 mL最大容量。作者提示临床强度磁场可致装置自动扩张,不应为携带AeroForm扩张器的女性开具磁共振检查。

要点

  • AeroForm扩张器在临床强度磁场下可能发生自动充气扩张。
  • 本例双侧扩张器在磁共振检查室突然激活并充至800 mL容量。
  • 携带AeroForm扩张器者应避免磁共振检查。

核心结论

  • V2.5 AeroForm扩张器过度充气率虽低于前代,但自动扩张仍是需关注的问题。
  • 装置在临床强度磁场下可能通过不同机制发生自动扩张。
  • 不应为携带AeroForm扩张器的女性开具磁共振检查。

研究设计与样本

  • 个案报道:一例57岁女性stage 1导管原位癌(T1c=1.3 cm,ER/PR阴性,HER2阳性)。
  • 行双侧保留乳头乳房切除、前哨淋巴结活检及V2.5 AeroForm扩张器重建。

主要发现

  • 术后及随访初期无并发症。
  • 进入磁共振检查室评估偶然发现的脑膜瘤时,双侧扩张器二氧化碳储气阀突然激活。
  • 气体释放使装置充至800 mL最大容量。
  • 患者选择保留扩张器,1个月后更换为永久硅胶假体。

临床怎么用

  • 携带AeroForm扩张器的女性不应安排磁共振检查。
  • 临床需警惕磁场环境下扩张器自动充气的风险。

局限性

  • 本文为单一个案报道,不能确定发生率或因果关系。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Device-based reconstruction, the most common approach to breast reconstruction, typically occurs in 2 stages to allow for the breast envelope to be expanded serially before exchanging for a permanent implant. The AeroForm tissue expander is a carbon dioxide-based expansion system that allows for patient-initiated filling, and compared with the traditional saline expander, it has been shown to decrease the median time to full expansion and implant exchange. Furthermore, multiple changes have been incorporated into the newer V2.5 iteration resulting in reductions in device-related malfunctions, including overinflation. CASE PRESENTATION: A 57-year-old woman who had a diagnosis of stage 1 ductal carcinoma in situ (T1c = 1.3 cm, ER/PR-negative, HER2-positive) underwent bilateral nipple-sparing mastectomy with sentinel lymph node biopsy and V2.5 AeroForm expander reconstruction. The procedure and subsequent postoperative follow-up visits were unremarkable without complications. However, once she entered an magnetic resonance imaging examination room for evaluation of an incidentally found meningioma, the carbon dioxide reservoir valve in both expanders was suddenly activated, releasing filling gas and maximally inflating the device to its 800-mL capacity. She elected to retain the expanders, and they were exchanged for permanent silicone implants 1 month later. CONCLUSIONS: Although the V2.5 AeroForm expander is associated with a much lower rate of overinflation than its predecessor, our case suggests that autoexpansion remains of concern and likely occurs via a different mechanism when the device is subjected to clinical strength magnetic field. Clinicians should not order magnetic resonance imaging scans for women with AeroForm expanders due to the risk of unintended expansion.

来源:Ann Plast Surg,查看原文。版权归原作者所有。

引用

  1. Huang H, Bogue JT, Otterburn DM, et al. AeroForm Tissue Expander: A Case of Early Accidental Inflation. Ann Plast Surg. 2021. PMID: 33538501. doi:10.1097/SAP.0000000000002678

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。