微血管吻合器在动脉端端吻合中的应用:系统综述
该系统综述检索4个数据库至2020年11月,纳入20项研究共1639例患者,涉及670个动脉和1124个静脉吻合。动脉吻合器总成功率为92.1%(617/670),8%出现故障或失败,多发生于肢体重建。动脉耦合应用不广泛,主要集中于头颈和胸部重建。
要点
- 动脉吻合器总成功率为92.1%(617/670)。
- 8%动脉吻合出现故障或失败,多见于肢体重建。
- 应用集中于头颈和胸部重建,因动脉壁厚或钙化而受限。
核心结论
- 动脉耦合应用不广泛,主要集中于头颈和胸部重建,总报告有效率为92.1%。
研究设计与样本
- 系统综述,检索4个电子数据库至2020年11月,遵循PRISMA指南,最终纳入20项研究。
主要发现
- 共1639例患者,行670个动脉吻合和1124个静脉吻合。
- 动脉吻合中351个用于头颈游离组织移植,117个用于乳房重建,4个用于上肢重建,5个用于下肢重建,其余未说明。
- 动脉吻合器总成功率为92.1%(617/670)。
临床怎么用
- 显微外科医生因动脉固有特征(壁厚、钙化、几何差异)不常规使用现有吻合器装置。
- 研究提供了微血管动脉耦合的集体建议及「应做与不应做」事项。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: There are several reasons microsurgeons may not use a coupler device in arterial anastomosis: may be thick-walled, non-pliable due to atherosclerotic calcification or present vessel geometrical discrepancies. This review summarises the current applications, efficacy and troubleshooting of microvascular coupler devices in arterial end-to-end anastomosis. METHODS: A systematic review of the literature was performed in November 2020 across 4 electronic databases and in accordance with the PRISMA guidelines. All studies comprised the data synthesis that reported the use of a microvascular coupler device for arterial end-to-end anastomosis. Data were extracted and collected in three groups of standardised variables: study, anastomosis-related and technical characteristics. RESULTS: Out of the 7,690 articles identified, 20 were included in the final data synthesis. Included studies involved a total of 1639 patients, who underwent 670 arterial and 1,124 venous anastomoses. Out of all arterial anastomoses, 351 were performed in free tissue transfers in head and neck, 117 in breast, 4 in upper extremity and 5 in lower extremity reconstruction, whereas the remaining were not specified. The total arterial coupler anastomosis success rate reported was 92.1% (617/670). Fifty-three (8%) arterial anastomoses were reported to result in either troubleshooting events or intra- or post-operative failures, most being reported in extremity reconstructions. CONCLUSIONS: Arterial coupling is not widespread with predominant use in head and neck and chest reconstructions, and total reported efficacy of 92.1%. Microsurgeons are reluctant to routinely use current widespread coupler devices as a result of inherent arterial characteristics. This study delivered collective recommendations, 'do's and don'ts' of microvascular arterial coupling.
引用
- Pafitanis G, Nicolaides M, O'Connor EF, et al. Microvascular anastomotic arterial coupling: A systematic review. J Plast Reconstr Aesthet Surg. 2021. PMID: 33551362. doi:10.1016/j.bjps.2020.12.090
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。