文献速递乳房整形与再造显微修复重建

PEAK PlasmaBlade™与传统电刀在腹部游离皮瓣乳房重建中的随机对照试验

Comparison of PEAK PlasmaBlade™ to conventional diathermy in abdominal-based free-flap breast reconstruction surgery-A single-centre double-blinded randomised controlled trial.
PubMed 文献 · J Plast Reconstr Aesthet Surg · 2020-12-13

单中心双盲随机对照试验纳入108例腹部游离皮瓣乳房重建患者,比较PEAK PlasmaBlade™(PPB,n=56)与传统电刀(n=52)行皮瓣切取。主要结局腹部引流天数无显著差异(p=0.48),引流总量、炎性细胞因子及并发症亦无差异;2周随访时PPB组腹部超声下血清肿有减少趋势(p=0.09)且体积显著更小(p=0.04)。

要点

  • PPB未显著减少腹部引流天数、引流总量或炎性细胞因子。
  • 2周随访PPB组血清肿有减少趋势且体积显著更小。
  • 两组并发症发生率无显著差异。

核心结论

  • 使用PPB未显著减少引流需求、引流总量或炎性细胞因子,但减小了2周随访时血清肿体积,可能减少拔管后早期血清肿形成。

研究设计与样本

  • 单中心双盲随机对照试验,纳入108例腹部游离皮瓣乳房重建患者,PPB组n=56,传统电刀组n=52。
  • 主要结局为腹部引流所需天数,数据收集涵盖住院及门诊随访。

主要发现

  • 基线特征除PPB组皮瓣重量显著更低外其余相似。
  • 引流所需天数中位数无显著差异(p=0.48;传统电刀组6.0天,PPB组5.0天)。
  • 引流总量(p=0.68)、引流液炎性细胞因子(p>0.054)及并发症(p>0.24)两组无显著差异。
  • 2周随访时PPB组腹部超声下血清肿有减少趋势(p=0.09)且体积显著更小(p=0.04)。

临床怎么用

  • PPB可作为腹部皮瓣切取的备选设备,但不应期望缩短引流时间或减少引流总量。
  • 对关注拔管后早期血清肿形成的患者,PPB可能有一定优势,需结合2周超声随访评估。

局限性

  • 单中心研究,样本量108例,外推性有限。
  • PPB组皮瓣重量显著低于传统电刀组,可能影响组间可比性。
  • 血清肿减少仅为2周随访的趋势性发现,需更长随访验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Electrosurgery makes dissection with simultaneous haemostasis possible. The produced heat can cause injury to the surrounding tissue. The PEAK PlasmaBlade™(PPB) is a new electrosurgery device which may overcome this by having the ability to operate on a lower temperature, therefore reducing collateral thermal damage. METHOD: A single-centre, double-blinded, randomised controlled trial (RCT) was conducted which included 108 abdominal-based free-flap breast reconstruction patients who had their flap raise performed using either the PPB (n = 56) or the conventional diathermy (n = 52). Data were collected during their in-patient stay and out-patient appointments. The primary outcome value was the number of days the abdominal drains were required. RESULTS: Baseline characteristics were similar between the groups, except a significantly lower flap weight in the PPB group. The median number of days the drains were required did not differ significantly (p = 0.48; 6.0 days for the diathermy and 5.0 days for the PPB). The total drain output (p = 0.68), the inflammatory cytokine in the drain fluid (p>0.054) and complications (p>0.24) did not differ significantly between the two groups. At the 2-week follow-up appointment, there was a trend towards less abdominal seromas on abdominal ultrasound (p = 0.09) in the PPB group which were significantly smaller (p = 0.04). CONCLUSION: The use of the PPB did not result in a significant reduction of drain requirement, total drain output or inflammatory cytokines but did reduce the size of seroma collections at the 2-week follow-up appointment. Therefore, the use of the PPB device could reduce early seroma formation after drain removal.

来源:J Plast Reconstr Aesthet Surg,查看原文。版权归原作者所有。

引用

  1. Friebel TR, Narayan N, Ramakrishnan V, et al. Comparison of PEAK PlasmaBlade™ to conventional diathermy in abdominal-based free-flap breast reconstruction surgery-A single-centre double-blinded randomised controlled trial. J Plast Reconstr Aesthet Surg. 2020. PMID: 33422499. doi:10.1016/j.bjps.2020.12.007

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。