烧伤重症监护中自动床旁血糖监测系统的多中心可行性研究结果

Results of a Multicenter Feasibility Study of an Automated Bedside Glucose Monitoring System in the Burn Intensive Care Setting.
PubMed 文献 · J Burn Care Res · 2020-01-01

结果显示97.2%的Optiscanner结果落在YSI对应值25%以内,100%落在30%以内,平均绝对相对误差为9.6%,提示该连续血糖监测仪在危重烧伤患者中可提供准确的血糖测量。

要点

  • Optiscanner连续血糖监测在危重烧伤患者中与YSI测量一致性良好。
  • 97.2%的结果在YSI对应值25%以内,100%在30%以内。
  • 平均绝对相对误差为9.6%,支持进一步临床验证。

核心结论

  • 具有在线能力的连续血糖监测仪在危重烧伤患者中可提供准确的血糖测量。

主要发现

  • 97.2%的Optiscanner结果落在对应Yellow Springs Instrument值的25%以内,100%落在30%以内。
  • 平均绝对相对误差计算为9.6%。

关键数值与推荐

  • 97.2%的Optiscanner结果在YSI对应值25%以内。
  • 100%的Optiscanner结果在YSI对应值30%以内。
  • 平均绝对相对误差为9.6%。

临床怎么用

  • 该连续血糖监测仪有潜力提高烧伤重症监护中血糖监测频率,但治疗仍应参照现有医院方案使用单独获取的血糖值指导。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Intensive blood glucose regimens required for tight glycemic control in critically ill burn patients carry risk of hypoglycemia and are ultimately limited by the frequency of which serum glucose measurements can be feasibly monitored. Continuous inline glucose monitoring has the potential to significantly increase the frequency of serum glucose measurement. The objective of this study was to assess the accuracy of a continuous glucose monitor with inline capability (Optiscanner) in the burn intensive care setting. A multicenter, observational study was conducted at two academic burn centers. One hundred and six paired blood samples were collected from 10 patients and measured on the Optiscanner and the Yellow Springs Instrument. Values were plotted on a Clarke Error Grid and mean absolute relative difference calculated. Treatment was guided by existing hospital protocols using separately obtained values. 97.2% of results obtained from Optiscanner were within 25% of corresponding Yellow Springs Instrument values and 100% were within 30%. Mean absolute relative difference was calculated at 9.6%. Our findings suggest that a continuous glucose monitor with inline capability provides accurate blood glucose measurements among critically ill burn patients.

来源:J Burn Care Res,查看原文。版权归原作者所有。

引用

  1. Elder CT, Thigpin T, Karlnoski R, et al. Results of a Multicenter Feasibility Study of an Automated Bedside Glucose Monitoring System in the Burn Intensive Care Setting. J Burn Care Res. 2020. PMID: 31633750. doi:10.1093/jbcr/irz171

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。