Pliance X测量儿童压力衣与烧伤瘢痕界面压力的信度与临床实用性
该研究基于试点纵向队列数据,评估Pliance X在烧伤后接受压力衣治疗的儿童中测量界面压力的重测信度与临床实用性。
要点
- Pliance X测量儿童压力衣下静态界面压力的重测信度ICC=0.87(n=27),接近临床可接受水平。
- 设备对未哭闹的各年龄段儿童具有适用性和可接受性。
- 仍需进一步信度测试,包括评分者间信度,并可通过改良传感器和即时平均压力读数提升临床实用性。
核心结论
- Pliance X在儿童烧伤后压力衣治疗中测量静态界面压力的重测信度接近临床可接受水平,临床实用性对未哭闹的各年龄段儿童可接受。
研究设计与样本
- 数据来自一项试点纵向队列研究,基线时评估即时重测信度,采用组内相关系数(ICC)分析,并以适用、可及、可行和可接受四项标准评估临床实用性。
主要发现
- 静态界面压力的重测信度ICC=0.87(n=27),接近临床可接受水平。
- Pliance X对未哭闹的各年龄段儿童具有适用性和可接受性。
关键数值与推荐
- 静态界面压力重测信度ICC=0.87(n=27)。
- 临床实用性评估采用适用、可及、可行和可接受四项标准。
临床怎么用
- Pliance X可用于测量儿童压力衣与烧伤瘢痕间的静态界面压力,但需注意重测信度仅接近临床可接受水平。
- 对哭闹儿童的使用受限,建议在儿童未哭闹时进行测量。
局限性
- 需进一步开展信度测试,包括评分者间信度。
- 设备的临床实用性可通过替代传感器样式和即时平均压力读数来改善。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Reliable, clinically acceptable pressure measuring devices are important to accurately record interface pressure. The Pliance X is a device that may overcome previous device limitations (i.e. pressure garment distortion, temperature dependent sensors). This research aimed to identify the test-retest reliability and clinical utility of the Pliance X in children receiving pressure garment therapy post-burn. METHODS: Data was collected as part of a pilot longitudinal cohort study. Immediate test-retest reliability was investigated at baseline and analysed using Intra-Class Correlation Coefficients (ICCs). Clinical utility was evaluated at baseline using the criteria of appropriate, accessible, practicable and acceptable. RESULTS: Thirty-one children completed baseline measurements. Participants had a median age of three years (IQR: 1.5, 7.5), and 61% were male. Test-retest reliability of the Pliance X was close to acceptable for clinical use for stationary interface pressure (ICC=0.87, n=27). The Pliance X was appropriate and acceptable for children of all ages who were not distressed. DISCUSSION: Further reliability testing is required, including inter-rater reliability testing. The device's clinical utility could be improved with alternate sensor styles and an instant mean pressure reading to enable immediately clinically useful information on the interface pressure.
引用
- Wiseman J, Simons M, Kimble R, et al. Reliability and clinical utility of the Pliance X for measuring pressure at the interface of pressure garments and burn scars in children. Burns. 2018. PMID: 30077488. doi:10.1016/j.burns.2018.05.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。