文献速递颅颌面外科

术中导航能否提高下颌角截骨术的准确性:增强现实导航、个性化导板与徒手技术的比较

Does intraoperative navigation improve the accuracy of mandibular angle osteotomy: Comparison between augmented reality navigation, individualised templates and free-hand techniques.
PubMed 文献 · J Plast Reconstr Aesthet Surg · 2018-04-11

该回顾性研究纳入2011年9月至2016年6月收治的93例下颌角肥大术后患者,比较增强现实导航组(31例)、个性化导板组(28例)与徒手组(34例)的截骨准确性。结果显示徒手组准备时间显著短于增强现实导航组和个性化导板组(P<0.01),三组总手术时间无显著差异(P>0.05),增强现实导航组和个性化导板组截骨线偏差均低于徒手组(P<0.01)。

要点

  • 徒手组准备时间显著短于增强现实导航组和个性化导板组(P<0.01)。
  • 三组总手术时间无显著差异(P>0.05)。
  • 增强现实导航组和个性化导板组截骨线偏差均低于徒手组(P<0.01)。

核心结论

  • 基于增强现实(AR)的导航系统在某些特定临床操作中比其他技术具有更高的准确性、可靠性和用户友好性,具有广阔的临床前景。

研究设计与样本

  • 回顾性研究,纳入2011年9月至2016年6月于整形与重建外科收治的93例下颌角肥大术后患者。
  • 增强现实导航组31例,个性化导板组28例,徒手组34例。
  • 术后CT图像与三维软件生成的术前规划进行对比分析,测量准备时间、截骨时间、总手术时间和截骨线偏差。

主要发现

  • 准备时间徒手组显著短于增强现实导航组和个性化导板组(P<0.01)。
  • 总手术时间三组间无显著差异(P>0.05)。
  • 截骨线偏差增强现实导航组和个性化导板组均低于徒手组(P<0.01)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Augmented reality (AR)-based navigation surgery has evolved to be an advanced assisted technology. The aim of this study is to manifest the accuracy of AR navigation for the intraoperative mandibular angle osteotomy by comparing the navigation with other interventional techniques. METHODS: A retrospective study was conducted with 93 post-surgical patients with mandibular angle hypertrophy admitted at our plastic and reconstructive surgery department between September 2011 and June 2016. Thirty-one patients received osteotomy conducted using a navigation system based on augmented reality (AR group), 28 patients received osteotomy conducted using individualised templates (IT group) and the remaining 34 patients received osteotomy performed by free hand (free-hand group). The post-operative computed tomography (CT) images were reviewed and analysed by comparing with pre-surgical planning generated by three-dimensional (3D) software. The preparation time, cutting time, whole operating time and discrepancy in osteotomy lines were measured. RESULTS: The preparation time was much shorter for the free-hand group than that for the AR group and the IT group (P < 0.01). However, no significant difference in the whole operating time was observed among the three groups (P > 0.05). In addition, the discrepancy in osteotomy lines was lower for the AR group and in the IT group than for the free-hand group (P < 0.01). CONCLUSION: The navigation system based on AR has a higher accuracy, more reliability and better user friendliness for some particular clinical procedures than for other techniques, which has a promising clinical prospect.

来源:J Plast Reconstr Aesthet Surg,查看原文。版权归原作者所有。

引用

  1. Zhu M, Liu F, Zhou C, et al. Does intraoperative navigation improve the accuracy of mandibular angle osteotomy: Comparison between augmented reality navigation, individualised templates and free-hand techniques. J Plast Reconstr Aesthet Surg. 2018. PMID: 29729839. doi:10.1016/j.bjps.2018.03.018

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。