新型三维立体摄影测量工具用于手臂体积测量的准确性与可重复性
本研究在10名健康受试者腕横纹近端使用自主研发工具放置人工标记点,通过三维立体摄影获取手与前臂图像并测量体积,重复测量评估标记点放置导致的组内与组间变异。手部体积组内与组间ICC分别为0.96和0.98,前臂均为0.99,表明该方法高度可重复、快速且简便,能更细致分析淋巴水肿的定位。
要点
- 使用人工标记点的三维立体摄影测量手与前臂体积高度可重复。
- 手部体积组内ICC=0.96、组间ICC=0.98;前臂组内与组间ICC均=0.99。
- 该工具可更细致分析淋巴水肿的定位。
核心结论
- 使用人工标记点测量手与前臂体积被认定为高度可重复、快速且简便的操作,该工具能够更细致地分析淋巴水肿的定位。
研究设计与样本
- 在10名健康受试者腕横纹近端使用自主研发的测量工具放置人工标记点,获取手与前臂三维图像并进行体积测量,重复测量以评估标记点放置导致的组内与组间变异。
主要发现
- 手部体积测量的组内ICC=0.96、组间ICC=0.98,前臂体积测量的组内ICC=0.99、组间ICC=0.99。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Lymph edema of the arm is a common complication after breast cancer treatment. To evaluate lymph edema volume and treatment outcome, an easy to use, objective quantification method of arm volume is necessary. Most often, water displacement is used to measure arm volume, as it is an easy and robust method that can be performed at any place with a simple equipment. However, when using water displacement, no exact localization of volume difference can be seen. To accurately measure hand and forearm volumes separately, an in-house-developed device was constructed for an accurate landmark placement. METHOD: An in-house developed measurement tool was used for placing artificial landmarks proximal to the wrist crease on ten healthy subjects. Three-dimensional (3D) images of the hand and forearm were acquired, and volume measurements of the hand and forearm were performed. Measurements were repeated to investigate the intra- and inter-rater variability caused by the landmark placement. RESULTS: Measuring volume of the hand and forearm while using artificially placed landmarks turned out to be a highly reproducible, quick, and easy procedure. Both intra- and inter-rater variability showed high reproducibility for hand (ICC = 0.96 and 0.98, respectively) and forearm (ICC = 0.99 and 0.99, respectively) volumes. CONCLUSION: Measuring volumes of the hand and forearm while using artificially placed landmarks was found to be a highly reproducible, quick, and easy procedure. The device enables to analyze the localization of lymph edema in greater details.
引用
- Verhulst AC, Wesselius TS, Glas HH, et al. Accuracy and reproducibility of a newly developed tool for volume measurements of the arm using 3D stereophotogrammetry. J Plast Reconstr Aesthet Surg. 2017. PMID: 28874308. doi:10.1016/j.bjps.2017.07.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。