可见光光谱与手持多普勒用于游离皮瓣术后监测的前瞻性对比试验
所有皮瓣平均组织氧饱和度为56.7±7.1%(范围39.4–72.1%),不随患者合并症或皮瓣类型显著变化。3个皮瓣因灌注异常重返手术室探查,挽救率100%;VLS的灵敏度、特异度和准确度均高于间歇多普勒和临床检查。
要点
- VLS可连续无创监测皮瓣组织氧饱和度,弥补间歇多普勒对静脉淤血不敏感的不足。
- 81个皮瓣平均组织氧饱和度56.7±7.1%,不随合并症或皮瓣类型显著变化。
- VLS的灵敏度、特异度和准确度均优于间歇手持多普勒和临床检查。
核心结论
- 可见光光谱是游离组织移植监测的可靠连续辅助手段,优于间歇手持多普勒和临床检查。
研究设计与样本
- 前瞻性对照研究,评估T-Stat Model 303 VLS血氧仪与间歇多普勒及临床检查的诊断效能。
主要发现
- 所有皮瓣平均组织氧饱和度为56.7±7.1%(范围39.4–72.1%),不随患者合并症或皮瓣类型显著变化。
- 研究期间3个皮瓣因灌注异常重返手术室探查,挽救率为100%。
- VLS的灵敏度、特异度和准确度均高于间歇多普勒和临床检查。
临床怎么用
- 可将VLS作为游离皮瓣术后连续监测的辅助手段,与临床检查联合使用以及时发现灌注异常。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Early detection of compromised free flap perfusion is critical. A common modality of thrombosis detection, physical examination augmented with hand-held Doppler, provides only intermittent data and is insensitive to venous compromise. Visible light spectroscopy provides continuous, noninvasive evaluation of tissue perfusion. The authors hypothesized that visible light spectroscopy is a more sensitive and specific monitoring method for early detection of postoperative flap compromise than intermittent hand-held Doppler and clinical examination. METHODS: The authors prospectively conducted a controlled study evaluating the sensitivity, specificity, and accuracy of the T-Stat Model 303 VLS oximeter versus that of intermittent Doppler and clinical examination. The authors prospectively collected and analyzed patient data, complications, reoperations, flap failures, and tissue oxygen saturation. RESULTS: Sixty-eight patients with 81 flaps completed the study. The majority of flaps (86.4 percent) were either transverse rectus abdominis musculocutaneous or deep inferior epigastric artery perforator flaps. The mean tissue oxygen saturation for all flaps in the study was 56.7 ± 7.1 percent (range, 39.4 to 72.1 percent) and did not differ significantly with patient comorbidity or flap type. During the course of the study, three flaps were returned to the operating room for exploration because of a perfusion abnormality, and the resulting salvage rate was 100 percent. The sensitivity, specificity, and accuracy of visible light spectroscopy were found to be greater than both intermittent Doppler and clinical examination. CONCLUSION: Visible light spectroscopy is a reliable, continuous adjunct to free tissue transfer monitoring with advantages over intermittent hand-held Doppler and clinical examination. CLINICAL QUESTION/LEVEL OF EVIDENCE: Diagnostic, II.
引用
- Mericli AF, Wren J, Garvey PB, et al. A Prospective Clinical Trial Comparing Visible Light Spectroscopy to Handheld Doppler for Postoperative Free Tissue Transfer Monitoring. Plast Reconstr Surg. 2017. PMID: 28582335. doi:10.1097/PRS.0000000000003600
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。