MAVIS III 三维立体摄影测量评估增生性瘢痕体积的准确性与可靠性
本研究评估便携式三维主动立体摄影测量系统 MAVIS III 测量烧伤后增生性瘢痕体积的准确性与评估者间可靠性。以可塑油泥模型对比液体置换法确定准确性,并由三名盲法评估者对25张不同解剖部位瘢痕图像进行体积测量以评估可靠性。结果显示体积测量准确性为充分至优秀(Pearson相关系数0.54-0.90),人体增生性瘢痕体积测定的总体组内相关系数为0.95,评价为优秀。
要点
- MAVIS III 测量增生性瘢痕体积与液体置换法相比准确性充分至优秀。
- 评估者间可靠性优秀,总体组内相关系数为0.95。
- 该设备有望用于烧伤后增生性瘢痕的无创体积评估。
核心结论
- MAVIS III 在体积测量方面准确且在不同使用者之间可靠,是增生性瘢痕无创评估的潜在候选工具。
研究设计与样本
- 采用可塑油泥制作不同皮肤与创面类型的增生性瘢痕模型,将相机软件计算的体积与液体置换法测得的体积进行比较以确定准确性。
- 由三名盲法评估者对25张涵盖手、臂、腿和胁部等不同解剖区域的增生性瘢痕患者图像进行评估,以评价评估者间可靠性。
主要发现
- MAVIS III 获得的体积测量与液体置换法测定相比,准确性为充分至优秀,Pearson相关系数为0.54-0.90。
- 人体增生性瘢痕体积测定的可靠性评价为优秀,总体组内相关系数为0.95。
临床怎么用
- 可将 MAVIS III 作为烧伤后增生性瘢痕体积的无创评估工具,用于不同评估者之间的体积测量。
- 研究提示需进一步优化主动立体摄影测量及软件用户界面,以提升瘢痕体积评估效果。
局限性
- 研究强调需进一步改进主动立体摄影测量和软件用户界面,以实现更优的瘢痕体积评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Current strategies for estimating positive scar volume following burns is limited to employing subjective scar scales. This study assesses the accuracy and reliability of Measurement of Area and Volume Instrument System (MAVIS) III, a portable 3D active stereophotogrammetric imaging system, for non-invasive assessment of hypertrophic scar volume and elucidates factors that leads to inaccurate volume determination. METHODS: To determine accuracy, hypertrophic scars arising in various skin and wound types were modeled using moldable putty substance. Volumetric measurements calculated by the camera's software were compared to measurements determined by fluid displacement. To assess inter-rater reliability, 25 images of patients with hypertrophic scars on various anatomical regions (hand, arm, leg, and flank area) were assessed by three blind evaluators. RESULTS: Volumetric measurements obtained from MAVIS III ranged from adequate to excellent accuracy (Pearson's correlation coefficient of 0.54-0.90) compared to measurements determined by fluid displacement. The reliability of volume determination of human hypertrophic scars was rated excellent (overall Intraclass correlation coefficient of 0.95). CONCLUSIONS: This study finds that MAVIS III is accurate in its volumetric measurements and is reliable between users; making it a potential candidate for non-invasive assessment of hypertrophic scars. Further considerations for refining active stereo-photogrammetry and the user interface of the software are also highlighted to allow superior evaluation of scar volume.
引用
- Su S, Sinha S, Gabriel V, et al. Evaluating accuracy and reliability of active stereophotogrammetry using MAVIS III Wound Camera for three-dimensional assessment of hypertrophic scars. Burns. 2017. PMID: 28363664. doi:10.1016/j.burns.2017.02.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。