文献速递颅颌面外科

四种三维扫描系统用于颅颌面成像的准确性比较

Comparison of three-dimensional scanner systems for craniomaxillofacial imaging.
PubMed 文献 · J Plast Reconstr Aesthet Surg · 2017-01-07

本研究以3dMDface系统为参照,用四种三维扫描仪采集八名成年志愿者头型,以2mm内数据百分比和均方根距离评估准确性。M4D Scan准确性最高(90%在2mm内,RMS 0.71mm±0.28mm),其次为Avanto MRI(86%,1.11mm±0.33mm)和Structure Sensor(80%,1.33mm±0.46mm)。3dMDface与M4D Scan可提供高质量结果,扫描仪选择需兼顾临床与技术需求。

要点

  • M4D Scan相对3dMDface准确性最高,90%数据在2mm内。
  • Avanto MRI准确性86%,Structure Sensor为80%。
  • M4D Scan与Structure Sensor精度分别为0.50±0.04mm和0.51±0.03mm。

核心结论

  • 3dMDface系统与M4D Scan可提供高质量的颅颌面三维成像结果。
  • 紧凑型手持系统未来可能更受欢迎。

研究设计与样本

  • 采集八名成年志愿者头型,比较四种三维扫描仪。
  • 以3dMDface系统重建为参照评估准确性,以RMS评估精度。

主要发现

  • M4D Scan准确性最高:90%数据在2mm内,RMS为0.71mm±0.28mm。
  • Avanto MRI准确性86%,RMS为1.11mm±0.33mm。
  • Structure Sensor准确性80%,RMS为1.33mm±0.46mm。
  • M4D Scan精度为0.50±0.04mm,Structure Sensor为0.51±0.03mm。

关键数值与推荐

  • 准确性定义为距3dMDface重建2mm范围内的数据百分比。
  • 精度以表面间均方根距离确定。

临床怎么用

  • 临床与技术需求共同决定扫描仪选择。
  • 追求高质量结果时可考虑3dMDface系统或M4D Scan。

局限性

  • 样本为八名成年志愿者,未涉及儿童或颅颌面畸形患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Two-dimensional photographs are the standard for assessing craniofacial surgery clinical outcomes despite lacking three-dimensional (3D) depth and shape. Therefore, 3D scanners have been gaining popularity in various fields of plastic and reconstructive surgery, including craniomaxillofacial surgery. Head shapes of eight adult volunteers were acquired using four 3D scanners: 1.5T Avanto MRI, Siemens; 3dMDface System, 3dMD Inc.; M4D Scan, Rodin4D; and Structure Sensor, Occipital Inc. Accuracy was evaluated as percentage of data within a range of 2 mm from the 3DMDface System reconstruction, by surface-to-surface root mean square (RMS) distances, and with facial distance maps. Precision was determined by RMS. Relative to the 3dMDface System, accuracy was the highest for M4D Scan (90% within 2 mm; RMS of 0.71 mm ± 0.28 mm), followed by Avanto MRI (86%; 1.11 mm ± 0.33 mm) and Structure Sensor (80%; 1.33 mm ± 0.46). M4D Scan and Structure Sensor precision were 0.50 ± 0.04 mm and 0.51 ± 0.03 mm, respectively. Clinical and technical requirements govern scanner choice; however, 3dMDface System and M4D Scan provide high-quality results. It is foreseeable that compact, handheld systems will become more popular in the near future.